Installation
1
LCR Meter
27
Connecting Test Head
Connecting the DUT when using a special test fixture
When taking measurements while using special test fixtures
with 7- mm terminals like the Keysight 16196A, follow these
steps to connect the test head, test fixture stand, and
3.5- mm- to- 7- mm adapter.
Step 1.
As shown in
(1), attach the test head to
the fixture stand with two screws. At this point, do not
tighten the screws completely (there should be some play).
Step 2.
As shown in
(2), align the
3.5- mm- to- 7- mm adapter with the hole in the test fixture
stand and gently insert the adapter.
Step 3.
As shown in
(3), tighten the connector nut
of the 3.5- mm connector with the provided wrench (for
3.5mm/SMA connector).
Figure 4
Connecting the test head to test fixture stand