What is the B2900 SMU?

An SMU combines the capabilities of a current source, 
a voltage source, a current meter, and a voltage meter 
along with the capability to switch easily between these 
various functions into a single instrument. This gives it 
the ability to evaluate the IV characteristics of devices 
across all four measurement quadrants without the need 
for any additional equipment.

In addition to its DC operation mode, B2901B/B2902B/
B2911B/B2912B also have the ability to perform pulsed
measurements in order to prevent device self-heating
from distorting the measurement results.

-

A

-

V

Voltage source

Current meter

Current source

Voltage meter

2-Wire and 4-Wire Connections

To simplify the connections, use 2-wire connection by
connecting the Force terminals only and opening the 
Sense terminals.

For low resistance measurements and high current
measurements, use 4-wire connection. Connecting 
the Force and Sense lines together at the terminal of the 
DUT minimizes the measurement error caused by the 
residual resistance of the test leads or cables. 

Grounded and Floating Measurements

With the default setting, Low Force 
terminal is connected to the chassis 
ground. However, it can be
disconnected from the ground for 
floating measurements. This setup is 
effective for differential voltage
measurements which usually require
two channels as shown in the 
Grounded 2 connection.

DUT

Grounded 1

DUT

Grounded 2

High Force

Low Force

DUT

Floating

SMU

High Force

Low Force

SMU

High Force

Low Force

SMU

High Force

Low Force

SMU

For the floating measurements, set the Low Terminal State to FLOATING on the Output Connection dialog box.
With the default setting, it is set to GROUNDED. The dialog box is opened by pressing the Config > Source > 
Connection function keys. With the FLOATING setting, the Low Force and Low Sense terminals can be 
connected to the maximum of ±250 V.
WARNING: Potentially hazardous voltages of up to ±250 V may be present at the Low Force and Low Sense 
terminals. To prevent electrical shock, use accessories that comply with IEC 61010-031. The terminals and 
the extended conductors must be isolated by using insulation caps, sleeves, etc.

DUT

4-Wire

DUT

High Force

Low Force

SMU

2-Wire

High Force

Low Force

SMU

High Sense

Low Sense

Summary of Contents for B2900B Series

Page 1: ...ite To connect the device under test DUT optional 1 B2900 terminals use banana jacks Prepare test leads that use banana plugs 2 Connect the test leads as shown in the figures For using 4 wire connecti...

Page 2: ...urns the channel off if it is in the output status even if it is in the remote status Two switches on 2 channel models The switch turns green if the channel is enabled The switch turns red if the chan...

Page 3: ...e Ch2 B2902B Graph G Single 1 Roll B2911B Trigger auto Trigger active High voltage Floating Remote Local lockout Error MOVE EDIT LXI LAN status indicator View mode 1 2 3 4 MOVE status EDIT status 1 Pr...

Page 4: ...ly Performing Measurement Disabling the Channel Press the Ch1 switch to disable channel 1 This turns off the switch light Measured value a Press the View key to display the Single view Press to start...

Page 5: ...and the status changes to MOVE blue e Use the rotary knob arrow keys or numeric alpha keys to set the Start sweep start Stop sweep stop and Step sweep step or Points number of sweep steps values 4 Pr...

Page 6: ...t result Pass or Fail Single View Range Setup Sweep Setup Pulse Setup Trigger Setup Source Volts Spot Measure Amps Source Amps Spot Measure Volts Measure Ohms Constant voltage output range Current mea...

Page 7: ...cursors 1 and 2 Second line Positions and distance e g t1 t2 t of X cursors 1 and 2 1 2 3 4 5 6 7 8 9 10 Graph View 1 Displays status ON or OFF on B2911B Displays status Ch 1 Ch 2 or OFF on B2912B Tw...

Page 8: ...surement error caused by the residual resistance of the test leads or cables Grounded and Floating Measurements With the default setting Low Force terminal is connected to the chassis ground However i...

Reviews: