M9202A Front Panel Features
M9202A Front Panel Features
Front Panel Connectors
Connector
Description
TRG 1
This MMCX female connector is a 50
Ω terminated input for an external trigger sig-
nal. The trigger signal range is ± 5 V
TRG 2
This MMCX female connector is a 50 Ω terminated input for an external trigger sig-
nal. The trigger signal range is ± 5 V.
TRG 3
This MMCX female connector is a 50 Ω terminated input for an external trigger sig-
nal. The trigger signal range is ± 5 V.
INPUT 1
This SMA female connector is the analog signal input. This input is AC-coupled and
50
Ω terminated. The input full scale range is 1 V (+4
dBm). Frequency range: 30
MHz to 1 GHz.
CLK
1 IN
This SMA female connector is the external clock signal input. This input is AC-cou-
pled and 50
Ω terminated. The recommended signal level is 1 V pk-pk (+4
dBm). The
clock signal frequency should be half of the desired sampling frequency.
TRG OUT
This MMCX female connector features a trigger signal output.
REF
IN
This MMCX female connector is the clock external reference input (100 MHz). This
input is AC-coupled and 50 Ω terminated. The recommended signal level is 1 V pkpk
(+4 dBm).
Front Panel LED
LED
Description
STATUS
(off)
Power is not present or the module has a fault.
(green) Indicates that the trigger is armed and waiting for a valid trigger to
occur.
(red)
Indicates that a trigger has occurred, and the acquisition is complete.
(yellow)
The power-on state immediately after power is applied.
Also identifies the module associated with the current channel in the
Soft Front Panel when the 'Identify' function is used.
Agilent M9202A PXI Express 12-bit Wideband IF Digitizer Startup Guide
13