3-17
Use, Maintenance, and Care of the Devices
Connections
connector damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be
possible to do this by hand if one of the connectors is fixed (as on a test port). In all situations,
however, it is recommended that you use an open-end wrench to keep the body of the device
from turning. Refer to
for part number and ordering information.
1. Position both wrenches within 90 degrees of each other before applying force. See
. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting
action which can misalign and stress the connections of the devices involved. This is
especially true when several devices are connected together.
Figure 3-7 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
.
Figure 3-8 Using the Torque Wrench
Summary of Contents for 85054B
Page 1: ...Keysight Technologies 85054B 50 Type N Calibration Kit User s and Service Guide...
Page 2: ......
Page 3: ...85054B...
Page 7: ...1 General Information...
Page 15: ...2 Specifications...
Page 20: ...2 6 85054B Specifications Electrical Specifications...
Page 21: ...3 Use Maintenance and Care of the Devices...
Page 43: ...4 Performance Verification...
Page 46: ...4 4 85054B Performance Verification Recertification...
Page 47: ...5 Troubleshooting...
Page 52: ...5 6 85054B Troubleshooting Contacting Keysight...
Page 53: ...6 Replaceable Parts...
Page 58: ...6 6 85054B Replaceable Parts Introduction...
Page 59: ...A Standard Definitions...
Page 64: ...Index 85054B Index 4 part number 6 2 Z zeroing connector gage 3 8 3 10 3 12...