85054B
2-5
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications in
apply to the devices in your calibration kit when
connected with a Keysight precision interface.
Certification
Keysight Technologies certifies that this product met its published specifications at the time of
shipment from the factory. Keysight further certifies that its calibration measurements are
traceable to the United States National Institute of Standards and Technology (NIST) to the
extent allowed by the Institute's calibration facility, and to the calibration facilities of other
International Standards Organization members. See
“How Keysight Verifies the Devices in This
for more information.
Table 2-3 Electrical Specifications
Device
Frequency (GHz)
Parameter
Specification
Lowband loads
DC to
2
Return Loss
48
d
0.00398
)
Sliding loads
a
a. Assuming proper usage, the specifications for the residual return loss after calibration for the
sliding load termination include:
• the quality of the airline portions within the sliding load, combined with
• the effective stability of the sliding element.
Proper usage includes the following practices:
•
Connector mating surfaces are clean.
• The changes in slide positioning are NOT done in equal steps since this results in very poor
calibration for some portions of the frequency range.
• The center conductor of test port connectors are nominally set back from the outer
conductor.
Sliding loads are designed to allow the center conductor to be moved. The position of the
sliding load center conductor should be set by a reference block and not positioned flush
against the center conductor of the test port.
> 2 to
18
Return Loss
42
d
0.00794
)
Adapters
(both styles)
DC to
8
>
8 to
18
Return Loss
Return Loss
34
d
0.0200
)
28
d
0.0398
)
Offset Opens
b
b. The specifications for the opens and shorts are given as allowed deviation from the nominal
model as defined in the standard definitions (see
“Class Assignments and Standard
Definitions Values are Available on the Web” on page A-2
).
at 18
Deviation from Nominal
Phase
1.5
Offset Shorts
at 18
Deviation from Nominal
Phase
1.0
Summary of Contents for 85054B
Page 1: ...Keysight Technologies 85054B 50 Type N Calibration Kit User s and Service Guide...
Page 2: ......
Page 3: ...85054B...
Page 7: ...1 General Information...
Page 15: ...2 Specifications...
Page 20: ...2 6 85054B Specifications Electrical Specifications...
Page 21: ...3 Use Maintenance and Care of the Devices...
Page 43: ...4 Performance Verification...
Page 46: ...4 4 85054B Performance Verification Recertification...
Page 47: ...5 Troubleshooting...
Page 52: ...5 6 85054B Troubleshooting Contacting Keysight...
Page 53: ...6 Replaceable Parts...
Page 58: ...6 6 85054B Replaceable Parts Introduction...
Page 59: ...A Standard Definitions...