22
7. Measuring Procedure
(1)Select & Connect the probe
In the LZ-330J, select the appropriate type of
probe for the type of material to be measured.
(
P. 6,13)
(2) Turn on the main unit's power.
Press the ON key.
(3) Adjustment
Before beginning a measurement, check to see if
the unit has been calibration for measurement of
that type of material. If calibration has not been
performed, perform the calibration operation and
register the calibration settings. (
P. 15)
(4) Delete contents of the measurement
value memory
Except in cases in which you wish to continue
making measurements with data acquired previ-
ously, you must delete the entire contents of the
measurement value memory. (
P. 26)
(5) Connect the printer
Connect the printer to the unit if you wish to print
out results while performing measurement and
other operations. (It is also possible to connect
the unit to a printer after completing measure-
ments and print out the measurement values and
the results of statistical calculations.
(The printer is an optional accessory.)
(6) Specify calibration settings
Specify the calibration memory address of cali-
bration settings registered for the type of material
being measured. (
P. 15)
(Example: Specifying the calibration settings
stored at calibration memory address 2.)
• When the display appears as indicated below:
• When the display indicates a calibration memory address: