3) Standard Foil Calibration Procedure
Turn on the probe and the main unit and hold the tip of the probe toward the air for 2 ~ 3 seconds.
•
Example: Performing calibration using the zero foil and standard foils (40, 100 and 400
µ
m foils).
1
2
5
5
6
7
Press the
key.
Press the
key.
Measure the (substrate) 5 times. The buzzer will sound and
the measured value will be displayed each time.
• Substrate: An uncoated (no plating, paint, etc.) sample of the same shape
and material as that to be measured.
• Depending upon the type of material, the value displayed may differ greatly
from that indicated in the example to the left (a value close to zero).
However, the procedure can be continued by setting it close to the setting
value by performing steps
9
and
0
.
○
○○○
Step
Operation
Display
Explanation
(Measure 5 times)
<Calibration using the zero foil>