Step
Display
Operation
①
1 . C H
P A R A M E T E R
▶
2 . C H
D U P L I C A T E
3 . C H
R E S E T
4 . C O M
S E T T I N G
5 . P A S S
C O D E
6 . X P O R T
S E T U P
M O D E
● Select the channel parameter.
Move the cursor in the menu screen to [1. CH PAR AMETER]
with [
⬆
]/[
⬇
] key. Then, press the [ENTER] key to set.
②
1 . C H
P A R A M E T E R
0 1
F L O U R
1 2 3 4
▶
0 2
F L O U R 2
1 2
3
4
0 3
C O R N 1
1 2
− 4
0 4
C O R N 2
1
− − −
0 5
B U T T E R
1 2 3 −
0 6
− − − −
0 7
− − − −
● Select the channel to set.
Move the cursor to the desired channel with [
⬆
]/[
⬇
] key.
Then, press the [ENTER] key to select.
③
1 . C H
P A R A M E T E R
0 1
F L O U R
C H
N A M E
◀
F L O U R
▶
S M O O T H I N G
1
M E A S
N T
1
C O M P .
P A R A M E T E R
O K
▶
C A N C E L
▶
● Enter the channel name.
If the already entered, move the cursor to the right end [
▶
],
and press the [ENTER] key. The channel name will be set.
<To enter or edit>
Move the cursor using the [EXIT] and [ENTER] keys.
Input the channel name from the 1st character using the
numeric keypad moving the cursor one by one with the [Enter]
key. Up to 8 characters can be entered. After input is done,
move the cursor to the right end [
▶
] using the [ENTER] key.
Then, press the [ENTER] key again to set.
④
1 . C H
P A R A M E T E R
0 1
F L O U R
C H
N A M E
0
S M O O T H I N G
◀
1
▶
M E A S
N T
2
C O M P .
P A R A M E T E R
C A N C E L
▶
● Enter the smoothing value of continuous measurement.
Select the smoothing value (0 to 5) using [
⬆
]/[
⬇
] key, and
press the [ENTER] key to set.
A channel identification. It will
be displayed when changing
the channel, or displaying
measurement value.
1. Channel name (8 alphanumerics)
⇒
step (3)
2. Smoothing value of continuous
measurement (0 to 5)
It is set to minimize the variation of
the displayed value when observing
the time variation in the continuous
measurement mode or absorbance
display mode. If a large value is set
to the smoothing value of continuous
measurement, a smooth change without
quivering will be observed. However,
it will deteriorate the response.
Therefore, it should be avoided to set
unnecessarily large value.
⇒
step (4)
The averaging times in the normal
measurement mode can be set. The
default is "1 time". This unit count one
data by averaging 6 data. Therefore,
ke e p t h e d ef au l t i n t h e r e g u l a r
operation.It will help improve averaging
accuracy of the samples.
Measurement averaging times
<Once>
The required time for measurement is
approximately 9 seconds
<Twice>
The required time for measurement is
approximately 12 seconds
3. Averaging times in normal measurement
(Default: 1 time)
■ Setting the channel parameter
Following four items can be set per channel.
* To cancel the setting during operation, press the [EXIT] key. Then, the previous screen will be restored.
⇒
step (5)