16
8. Measurement
Measure the sample in the same measurement condition as making calibration curve. (→ 6. Setting
the measurement conditions (P.11))
* It is recommended to perform zero-adjustment. (→ 9. Zero-adjustment (P.18))
<Step 1>
● Select the channel number.
In the initial screen, select the channel operating [CH.], [
⬆
], [
⬇
], and [ENTER] keys.
[Channel selection screen]
When changing the measurement channel, or during setup, the following will be displayed.
0 1
F L O U R
1 2 3 4
▶
0 2
F L O U R 2
1 2
3
4
0 3
C O R N 1
1 2
− 4
0 4
C O R N 2
1
− − −
0 5
B U T T E R
1 2 3 −
0 6
− − − −
0 7
− − − −
0 8
− − − −
[Presence of coefficient of each composition and display setting]
Icon
Coefficient
Display setting
1234
Present
ON
1 2 3 4
Present
OFF
−
Absent
Setting not available
* When changing the measurement channel, the channel that has no coefficient or the one whose display
setting of all four compositions is set OFF cannot be selected.
<Step 2>
● Select the measurement mode.
Press the [MODE] key and select the measurement mode.
There are three modes: normal measurement mode, continuous measurement mode, and
absorbance display mode.
8-1. Normal measurement mode
Following step 1 and 2, select the normal measurement mode.
0 1
F L O U R
1
…
M O I S T U R E
2
…
P R O T E I N
1 3. 3
8. 8
3
…
C O M P 3
4
…
C O M P 4
1. 4
7 6. 1
Set the sample to cover up the measurement window and press the [MEA.] key. The composition
will be displayed after measurement.
* Set the number of display digits and the averaging times as required. (→ Setting the channel parameter (P.20))