3. If using File Holder A, remove File Holder A from the File
Holder Adapter and disconnect the File Holder Adapter from
the Probe Cord.
PRECAUTION
After using the device, refer to
“
Autoclavable Components:
Lip hook, File holder A and File holder B.
”
3
Turning the Device On/Off
Press the
[ ]
button on the left side of the device without
the file holder and lip hook touching anything. The device is
connected to power and turns on the display.
a. Once powered on, the calibration function will run
automatically.
[CAL]
flashes while the device is
being calibrated.
b. When calibration is complete, the measurement screen
appears on the display.
c. If the calibration function is not executed, Er will flash on
the measurement screen.
d. To end use, press and hold the
[ ]
button. The display
turns off.
PRECAUTION
If the device is powered on while the file holder and lip hook are
touching the object, the calibration function will not start, and
[Er]
is displayed on the measurement screen.
NOTE:
Calibration is completed in about 10 seconds, if the
measurement screen does not appear after 10 seconds,
contact Kerr Customer Care.
PRECAUTION
After using the device, refer to
“
Cleaning, Disinfection and
Sterilization Instructions
”
section to properly manage
the device.
PREPARING THE DEVICE
1
Preparation and Use Environment
The environmental information appropriate for using this device
is as follows.
•
Appropriate temperature: 10–35°C (50–95°F) / Appropriate
humidity: 30–75% / Atmosphere pressure: 700–1060 hPa
•
Do not use the device in a place that it is exposed to direct
sunlight, heat sources, or near a fire.
2
Assembly
•
Charge the battery sufficiently when using the device for the
first time or after long-term storage.
•
Be sure to charge the battery using the power cord and
AC/DC adapter included in the package.
•
Sterilize the lip hook and file holder when using the device
for the first time and after every patient.
•
Examine the role of each USB connector on the probe cord
and connect to the appropriate device.
•
Ensure the file holder and lip hook are properly connected
to the probe cord.
NOTE:
•
Refer to
“
Charging the Battery
”
on page 3 for more
information on how to charge and replace the battery.
•
Refer to
“
Assembling the Device
”
on page 3 for more
information on how to mount and remove the lip hook and
file holder.
USING THE DEVICE
1. Refer to
“
Initial Set-up
”
to connect the lip hook and file
holder to the device.
2. To turn on, press the
[ ]
button on the left side of the
device without the file holder and lip hook touching
anything.
PRECAUTION
If the device is powered on while the file holder and lip hook are
touching the object, the calibration function will not start, and
[Er]
is displayed on the measurement screen.
3. With the clip of the file holder B touching the lip hook,
ensure that the figures and graphs are activated on the
measurement screen.
4. While holding down the
[Push]
button on file holder B,
insert the file according to the shape of the clip.
PRECAUTIONS
• Use a stainless-steel hand file (k-file).
• Do not use deformed or damaged files.
5. Gently pull on the file to confirm it is secured.