en - ENGLISH

Apex Connect

ELECTRONIC APEX LOCATOR

CONTENTS OVERVIEW

 

1

  All Components

Apex Connect Unit

Probe Cord

File Holder Adapter

Lip Hook (5)

File Holder A

File Holder B

AC/DC Adapter

Power Cord

APEX CONNECT UNIT

No.

Name

Function

1

Power button

Turns on the main body. To turn off the power, press and hold the 

[ ]

 button for 2 seconds.

2

USB connector 

terminal

Use probe cord to connect the file holder and lip hook to the main body.

3

Display Screen

Displays the main body status and the measured value of the root canal length while the main body is in use.

4

Settings button

Press and hold the 

[ ]

 button to enter the settings mode. 

•  In the settings mode, the APEX can be set for the reference value, sound volume, and standby time 

5

Select button

In the settings mode, change the settings.

6

Battery cover

There is a built-in rechargeable battery that supplies power.

7

Charging terminal

Connects the power cord and AC/DC adapter to charge the battery.

LCD SCREEN

The measurement screen is the screen that appears on the display when the device is turned on. On this screen, the device can be checked and 

the root canal length read and set the usage environment.

No.

Description

1

Volume icon. The alert sound volume is displayed.

2

Battery Icon. The remaining battery charge is displayed.

3

Measurement information of the root canal length. The current location of the file in the root 

canal is displayed in a number and a graph.

Summary of Contents for Apex Connect

Page 1: ...Apex Connect INSTRUCTIONS FOR USE ELECTRONIC APEX LOCATOR...

Page 2: ...buttontoenterthesettingsmode Inthesettingsmode theAPEXcanbesetforthereferencevalue soundvolume andstandbytime 5 Selectbutton Inthesettingsmode changethesettings 6 Batterycover Thereisabuilt inrecharg...

Page 3: ...cause a malfunction Ensure conductors such as metal are not in contact with the metal part of the device There is a risk of device malfunction fire or electric shock Do not disassemble the device Dire...

Page 4: ...he display When charging is complete the battery icon stops blinking and stays on WARNING ChargeusingtheAC DCadapterandpowercordincludedin thepackage Besuretoconnectthepowerplugtoagroundedpowersupply...

Page 5: ...idity 30 75 Atmosphere pressure 700 1060 hPa Do not use the device in a place that it is exposed to direct sunlight heat sources or near a fire 2 Assembly Charge the battery sufficiently when using th...

Page 6: ...ed from the measuring device may be shown This is not a malfunction of the measuring device It may be due to the anatomical apex and or the physiologic apex may be located coronal to the radiographic...

Page 7: ...NOTE For more information on the components of the Elements Connect Cordless Endodontic Motor and how to use them refer to the Elements Connect Endodontic Motor Instructions for use 3 Assembly and Co...

Page 8: ...sasfollows R S Themotorstopsafterrotatingintheopposite directionofthecurrentsetting REV Themotorrotatesinthedirectionoppositetothe currentsetting Itthenrotatesinthesetdirectionagain 2 Itisthetorqueval...

Page 9: ...e of contamination 3 Rinse the lip hook in tap water for 30 seconds 4 Dry the lip hook with a soft cloth and dry at room temperature for at least 30 minutes 4 Packing and Sterilization NOTE Health Car...

Page 10: ...d errorsmayoccur so pleasestartthemeasurement withathickfiletoathinfile Thisisamulti rootedcanal tooth Checkwitharadiographifthere areanycracksintheteethand thenre measure Problem Cause Solution Erwil...

Page 11: ...15 ApexConnectFileHolderA 815 1816 ApexConnectFileHolderB 815 1817 ApexConnectReplacementBattery ELECTROMAGNETIC COMPATIBILITY This device has been tested for compliance with electromagnetic tests acc...

Page 12: ...errdental com Manufactured by Meta Systems Co Ltd 1214 18 Sicox tower 12F 484 Dunchon daero Jungwon gu Seongnam si Gyeonggi do 13229 Korea Distribution Facility Kerr Australia Pty Ltd 12 Mars Rd Unit...

Reviews: