TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
39
8.5 Changing the weight of the internal calibration weight
!
This change can be carried out solely by a specialist holding thorough knowledge of
handling scales.
!
Information on reference weights can be found on the Internet at
Display the menu item “Technical adjustment”, see chapter
To confirm, press and hold the
button until the sound signal stops.
Press and hold the
MENU
button until the sound signal stops. The scale will switch to
the weighing mode again automatically.
Ensure there are no objects on the scale plate. Press the
CAL
button.
Wait until the blinking weight value of the adjustment weight is displayed.
During
blinking, place the required adjustment weight carefully in the middle of the
scale plate. The blinking symbol will disappear.
Wait until the stabilization indicator is displayed, remove the adjustment weight.
Press and hold the
button until the sound signal stops.
The weight of the internal adjustment weight will be changed.
After the successful change, the scale will switch to the weighing mode again
automatically.