52
TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
11.4 Auto zero
This menu item enables to switch on or off the automatic zero point correction. In the
switched on state, the deviation or zero point disturbance is corrected automatically.
Tip:
If the amount of the weighed material is reduced or increased significantly, the
scale’s “stabilizing and compensating” mechanism can result in displaying erroneous
weighing results! (e.g.: slow outflow of the liquid from the container placed on the
scale, evaporating processes).
When dosing with small weight fluctuations, it is recommended to switch this function
off.
Using the navigation buttons
, select the menu item
<Auto zero>
.
Confirm by pressing the
button, the current setting will be displayed.
,
Using the navigation buttons
, select the required setting.
Symbol
Description
Auto zero off
“Auto zero” function is off
Auto zero 1
Range of the “Auto zero” function ±½ digits
Auto zero 2
Range of the “Auto zero” function ±3 digits
Auto zero 3
Range of the “Auto zero” function ±7 digits
Auto zero 3E
Range of the “Auto zero” function ±7 digits in the entire
weighing range
Confirm the selection, pressing the
button. The scale will be switched to
the menu again.
Return to the weighing mode, pressing the
ON/OFF