8.6 Adjustment report display/print
This function enables to print the report on the most recent adjustment.
The scale and printer communication parameters must be compatible.
Communication parameters, see chapter 11.2 and 11.3.
Printout conforming to GLP see chapter 12.8.
In the weighing mode press and hold the
MENU
button until the sound signal
stops. The configuration menu is displayed.
Using the navigation buttons
, select the menu item
<Calibration info>
.
Confirm, pressing the
button. The following will be displayed: date, time,
adjustment type and deviation when compared to the most recent adjustment.
After the optional printer is connected, the data can be printed by pressing the
button.
Sample printout (KERN YKB-01N):
27-08-20 10:41:17
Current date/time
Balance ID:
WI2000077
User ID
Miller
Project ID
KERN
Calibration mode
27-08-20 10:11:17 AM
Adjustment date/time
Internal calib.
Adjustment type
Corr.: 0.21 g
Deviation when compared
to the most recent adjustment
Signature:
Press the
ON/OFF
button. The scale will be switched to the menu again.
Whenever required, carry out other settings in the menu or press the
ON/OFF
button. The scale will switch to the weighing mode again.