TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
35
8.1 Adjustment mode selection
In the weighing mode press and hold the
MENU
button until the
sound signal stops. The configuration menu is displayed.
Using the navigation buttons
, select the menu item
<Calibration
mode>
.
Confirm by pressing the
button, the current setting will be
displayed.
Using the navigation buttons
, select the required setting.
Auto. calibration
Automatic adjustment using the internal weight.
Factory setting in the models in the verifiable setting.
Internal calib.
Adjustment using the internal weight after the
CAL
button
is pressed.
Not available in the models in the verifiable setting.
External calib.
Adjustment using the external weight, not available in the
models in the verifiable setting.
In models with the internal adjustment weight we do not
recommend adjustment using the external weight.
Technical calib.
Changing the weight of the internal calibration weight.
Not available in the models in the verifiable setting.
After the “Internal, external or automatic adjustment” option is
selected confirm by pressing the
button.
After the “Technical adjustment” option is selected, to confirm press
and hold the
button until the sound signal stops.
The scale will be switched to the menu again.
To leave the menu/return to the weighing mode, press and hold the
MENU
button until the sound signal stops.