26 Pipette calibration
Before starting the weighing process, acclimatise the sample liquid, weighing
container, pipette and pipette tip for at least 2 hours.
Use distilled water as a sample fluid.
•
Ambient temperature 20 – 25°C, temperature fluctuations during calibration
±
0.5°C
•
Relative humidity 50 -75 %
26.1 Settings <P2.13 PIPETTE CALIBRATION >
Applications can be adapted to your requirements by creating application-related
settings. The settings available depend on the currently enabled application.
A part of these settings is global, that is, they are – apart from a few exceptions –
identical to the weighing application (See chap 9.3 „Settings <P2.1 WEIGHING>”.
The description below merely deals with settings deviating from the default settings.
Navigation in menu see chap. 8.2
Menu item
Parameters
Explanation
P2.13.4
SHORTCUTS
Functions keys
F1; F2; F3; F4
For pipette calibration, the following additional function key is
available.
START
Start pipette calibration
P2.13.5
AMOUNT
TEST VOLUME
For devices with fixed volume select setting <1>
(test volume = nominal volume).
For devices with variable volume two or three volumes can
be tested:
1
As a rule 10 % of the nominal volume [µl]
2
As a rule 50 % of the nominal volume [µl]
3
As a rule 100 % of the nominal volume [µl]
P2.13.6
AMOUNT
MEASUREMENTS
Number of measurements per test volume,
selectable 6 -20 measurements
P2.13.7
TARE AUTO
YES
Automatic taring after test cycle
Turn on / off
NO:
26.2 Procedure
Summary of Contents for PLJ 300-3CM
Page 11: ...AEJ C AES C PLJ C BA d 1621 10 Abmessungen AEJ C AES C PLJ C...
Page 21: ...AEJ C AES C PLJ C BA d 1621 20...
Page 49: ...AEJ C AES C PLJ C BA d 1621 48...
Page 81: ...AEJ C AES C PLJ C BA d 1621 80 15 1 2 Benutzer l schen...
Page 83: ...AEJ C AES C PLJ C BA d 1621 82 15 2 1 Neues Produkt anlegen Produktbezeichnung festlegen...
Page 113: ...AEJ C AES C PLJ C BA d 1621 112...
Page 115: ...AEJ C AES C PLJ C BA d 1621 114...
Page 130: ...129 AEJ C AES C PLJ C BA d 1621 26 2 Durchf hrung Applikation anw hlen...
Page 173: ......
Page 182: ...10 AEJ C AES C PLJ C BA e 1621 Dimensions AEJ C AES C PLJ C...
Page 192: ...20 AEJ C AES C PLJ C BA e 1621...
Page 220: ...48 AEJ C AES C PLJ C BA e 1621...
Page 252: ...80 AEJ C AES C PLJ C BA e 1621 15 1 2 Delete user...
Page 254: ...82 AEJ C AES C PLJ C BA e 1621 15 2 1 Create new product Define product name...
Page 255: ...AEJ C AES C PLJ C BA e 1621 83 After entering a product name define all parameters one by one...
Page 284: ...112 AEJ C AES C PLJ C BA e 1621...
Page 286: ...114 AEJ C AES C PLJ C BA e 1621...
Page 301: ...AEJ C AES C PLJ C BA e 1621 129 Select application...
Page 352: ...10 AEJ C AES C PLJ C BA f 1621 Dimensions AEJ C AES C PLJ C...
Page 362: ...20 AEJ C AES C PLJ C BA f 1621...
Page 390: ...48 AEJ C AES C PLJ C BA f 1621...
Page 422: ...80 AEJ C AES C PLJ C BA f 1621 15 1 2 Effacement de l usager...
Page 454: ...112 AEJ C AES C PLJ C BA f 1621...
Page 456: ...114 AEJ C AES C PLJ C BA f 1621...
Page 471: ...AEJ C AES C PLJ C BA f 1621 129 26 2 R alisation S lectionner l application...