13 Adjustment < P1 JUSTAGE >
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Possible adjustments include:
•
Automatic adjustment using internal weight (models AEJ, PLJ)
time and temperature controlled, see chap. 13.1.
•
Manual internal adjustment (models AEJ, PLJ)
To activate, press the CAL-key, see chap. 13.2.1.
•
Adjustment using external weight (model AES), see chap. 13.2.2.
•
Observe stable environmental conditions. A warming up time (see chapter
1) is required for stabilization.
•
Internal adjustment can only be carried out when the weighing pan is free
of objects. If the weighing pan is loaded, the error message <Range
exceeded> will appear.
•
You can cancel the adjustment by pressing the
TARE-
key.
•
Carry out external adjustment as near as possible to the balance’s
maximum weight (recommended adjustment weight see chap. 1). Weights
of different nominal values (>30% max) may be used for adjustment but
are not optimal for technical
measuring. Info about test weights can be found
on the Internet at: http://www.kern-sohn.com.
Summary of Contents for PLJ 300-3CM
Page 11: ...AEJ C AES C PLJ C BA d 1621 10 Abmessungen AEJ C AES C PLJ C...
Page 21: ...AEJ C AES C PLJ C BA d 1621 20...
Page 49: ...AEJ C AES C PLJ C BA d 1621 48...
Page 81: ...AEJ C AES C PLJ C BA d 1621 80 15 1 2 Benutzer l schen...
Page 83: ...AEJ C AES C PLJ C BA d 1621 82 15 2 1 Neues Produkt anlegen Produktbezeichnung festlegen...
Page 113: ...AEJ C AES C PLJ C BA d 1621 112...
Page 115: ...AEJ C AES C PLJ C BA d 1621 114...
Page 130: ...129 AEJ C AES C PLJ C BA d 1621 26 2 Durchf hrung Applikation anw hlen...
Page 173: ......
Page 182: ...10 AEJ C AES C PLJ C BA e 1621 Dimensions AEJ C AES C PLJ C...
Page 192: ...20 AEJ C AES C PLJ C BA e 1621...
Page 220: ...48 AEJ C AES C PLJ C BA e 1621...
Page 252: ...80 AEJ C AES C PLJ C BA e 1621 15 1 2 Delete user...
Page 254: ...82 AEJ C AES C PLJ C BA e 1621 15 2 1 Create new product Define product name...
Page 255: ...AEJ C AES C PLJ C BA e 1621 83 After entering a product name define all parameters one by one...
Page 284: ...112 AEJ C AES C PLJ C BA e 1621...
Page 286: ...114 AEJ C AES C PLJ C BA e 1621...
Page 301: ...AEJ C AES C PLJ C BA e 1621 129 Select application...
Page 352: ...10 AEJ C AES C PLJ C BA f 1621 Dimensions AEJ C AES C PLJ C...
Page 362: ...20 AEJ C AES C PLJ C BA f 1621...
Page 390: ...48 AEJ C AES C PLJ C BA f 1621...
Page 422: ...80 AEJ C AES C PLJ C BA f 1621 15 1 2 Effacement de l usager...
Page 454: ...112 AEJ C AES C PLJ C BA f 1621...
Page 456: ...114 AEJ C AES C PLJ C BA f 1621...
Page 471: ...AEJ C AES C PLJ C BA f 1621 129 26 2 R alisation S lectionner l application...