PES/PEJ-BA-defsi-0510
103
6.8.2.
Bar graph display
In configuration menu 1 (chapter 7) you can activate/deactivate the bar graph dis-
play.
0
Hide bar graph
x B.G.
*
1
Show bar graph
Wägeplatte ist unbelastet
halbe Höchstlast
Waage ist mit Höchstlast belastet
The weighing range of the balance is divided into 40 graphic cuboids. Zero (0) will
appear on the graphic display if there is no weighing value on the balance. 20 graphic
cuboids are displayed if the balance is loaded up to one half of its weighing range.
6.8.3. Stability display
Stable
Unstable
If the display shows the stability display
[
]
the balance is in a stable status. The
[
]
indication disappears if the condition is unstable.
6.8.4. Balance zero display
Environmental influences can lead to the exact figure of “
000.0
“ not being displayed
in spite of an empty weighing dish. It is, however, possible to reset your balance to
zero at any time and thus ensure that weighing really does commence at zero. Set-
ting to zero when a weight is applied is only possible within a certain type-dependent
range. In the event that the balance cannot be reset to zero with an applied weight,
this range has been exceeded.
[
o - Err
]
will appear on the display.
If an exact zero reading is not displayed on the balance in spite of the weighing dish
being empty, press the TARE key and the balance will start resetting to zero. Your
balance will be set to zero after a short standby time.
In addition to this, the sign for the balance zero setting will be displayed
[
#
0
$
]
.
weihging dish unloaded
half of weighing range
weighing range is used
to full capacity
Summary of Contents for PEJ