PES/PEJ-BA-defsi-0510
108
EnD
CAL
!
a
z0
g
6.9.4. Adjustment test with internal weight (only PEJ)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required
for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Display
Activate function
[
w CA.
2
]
(see chap. 7).
w CA 2
!
a
z0
g
Start of the adjustment test:
The test is carried out automatically.
T. KnT
CAL
!
VAKT
!
T. 0
CAL
!
Summary of Contents for PEJ
Page 3: ......