MPD_M-BA-e-1620
15
6.3.1 Crucial features of performance
Note:
The personal scale MPD-M does not have any crucial features of
performance as per IEC 60601-1. The system may be subject to
interference by other devices even if these devices conform to current
emission requirements as per CISPR.
6.4 Minimum distances
Recommended safety distances between portable and mobile HF
telecommunication devices and the medical device
The medical device is designed for use in an electromagnetic environment in which HF disturbance
variables are controlled. The customer or user of the medical electrical device can help avoiding
electromagnetic disturbances by keeping the minimum distance between portable and mobile HF
telecommunication devices (transmitters) and the medical device – depending on the output
performance of the communication device, as stated below.
Rated capacity of
transmitter
W
The safety distance depends on the transmission frequency
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.20
1.20
2.30
10
3.80
3.80
7.30
100
12.00
12.00
23.00
For transmitters with a maximum rated capacity not stated in the table above you can calculate the
recommended safety distance in metres (m) yourself by using the equation belonging to each column,
whereby P equals the maximum rated capacity of the transmitter in Watt (W) as per details provided
by the transmitter manufacturer.
NOTE 1 Higher frequency range applies to 80 MHz and 800 MHz.
NOTE 2 These guidelines may not be applicable in all cases.
The spread of electromagnetic variables is influenced by absorption and reflections in
buildings, objects and humans.