KFB/KFN-TAM-BA_IA-e-1832
15
6.8 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
•
In weighing systems with a resolution of < 15 000 dividing steps an
adjustment is recommended.
In weighing systems with a resolution of > 15 000 dividing steps a
linearisation is recommended (see chap. 6.10).
•
Prepare the required adjustment weight. The weight to be used depends
on the capacity of the scale. Carry out adjustment as near as possible to
the scale’s maximum weight. Info about test weights can be found on
the Internet at: http://www.kern-sohn.com.
•
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
6.8.1 Verified weighing systems
In verified weighing systems the menu item for adjustment „P2 mode“ is
blocked.
KERN KFB-TAM
To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chap. 6.11
KERN KFN-TAM
To override the blocked access you will have to destroy the seal before calling
up the menu and to short-circuit the two contacts on the circuit board [K2],
using a jumper (See chap. 6.11).
Attention:
After destruction of the seal the weighing system must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can be
reused for applications subject to verification.
Summary of Contents for KFB-TAM
Page 67: ...KFB KFN TAM BA_IA d 1832 66 13 Anhang 13 1 Abmessungen Tischfuß Wandhalterung ...
Page 69: ......
Page 135: ...KFB KFN TAM BA_IA e 1832 67 13 Annex 13 1 Dimensions Support base wall bracket ...
Page 137: ......
Page 202: ...66 KFB KFN TAM BA_IA f 1832 13 Annex 13 1 Dimensions Pied de table support mural ...