20
KFB/KFN-TAM-BA_IA-e-1832
After linearisation the balance will carry out a self-test.
Remove adjusting weight
during
selftest, balance will return
into weighing mode automatically.
6.9.2
Non-verified weighing systems
Call-up menu item P3 CAL
Cal
Liner, see chap. 6.9.1
Confirm by
, the password query „Pn“ will be displayed.
Press
,
,
or
,
,
subsequently.
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press
.
When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait for
stability display, then press
.
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press
.
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait
for stability display, then press
.
After a successful linearisation the balance will carry out a
self-test. Remove adjusting weight
during
selftest, balance
will return into weighing mode automatically.
Summary of Contents for KFB-TAM
Page 67: ...KFB KFN TAM BA_IA d 1832 66 13 Anhang 13 1 Abmessungen Tischfuß Wandhalterung ...
Page 69: ......
Page 135: ...KFB KFN TAM BA_IA e 1832 67 13 Annex 13 1 Dimensions Support base wall bracket ...
Page 137: ......
Page 202: ...66 KFB KFN TAM BA_IA f 1832 13 Annex 13 1 Dimensions Pied de table support mural ...