EOB_M-BA-e-0710
16
Position of adjustment switch:
Position of adjustment switch Status
1. to the left
Not documented
2. concentric
adjustment position – enables adjustment
3. to the right
Verification position - Adjustment locked
6.9 Checking scale settings with respect to the calibration of weighing scales
For the adjustment function, the balance must be switched over to the service mode.
To achieve this, set the adjustment switch to adjustment position (See chpt. 6.8).
In the service mode all parameters of the balance can be modified. The service pa-
rameters may not be modified, as this could damage the balance settings.
6.9.1 Menu overview in service mode (adjustment switch in adjustment posi-
tion)
This overview exists merely for the examination of the set parameters by
the authorized Verification Standards Department.
Changes may only be made to the parameters of the automatic switch-off function
“
A.OFF
“ and the acoustic signal ”
BURR
“.
Navigation in the menu:
With the weighing scales switched on, press the
[
→
0
←
]
-key for c. 3 sec until
“SETUP“, followed by ”UNIT“ appears.
Press
the
[ TARE ]
-key repeatedly until the desired function appears.
Confirm the selected function by pressing the
[HOLD]
-key. The first parameter
appears. To select the desired parameter, use the
[HOLD]
-key and confirm by
pressing the
[TARE]
-key.
1 2 3