background image

OKN-1_OKO-1-BA-e-1510 

12 

4.1 

Reflected light unit

 

At  first  you  have  to  unite  the  lamp  housing  and  the  reflected  light  unit  on  their 
connection  points.  After  that  the  connection  has  to  be  fixated  by  an  Allen  screw, 
which is attached to the right-hand side of the connection point of the lamp housing. 
Now  the  slides  for  analyser,  polariser  and  colour  filter  can  be  attached  to  the 
appropriate slots (

see page 8

). 

Afterwards  in  order  to  install  the  reflected  light  unit  to  the  microscope,  you  must 
loosen the fixing screw on the tube connection point and remove the black protective 
cover. 
You can then insert the round dovetail bracket on the light unit into the round dovetail 
bracket on the housing and fix it with the fixing screw. When doing this, you should 
always make sure that you do not touch the lenses with your bare fingers and that no 
dust enters the apertures. 
Finally  you  must  establish  the  electrical  connection  between  power  supply  unit  and 
reflected light unit. The plug of the connection cable, attached to the rear of the light 
unit has to be installed to the socket on the rear of the power supply unit. Please do 
also use the screw lock on the plug.

  

4.2  Microscope head 

First  you  must  loosen the fixing  screw  on  the  tube  connection  point  of  the  reflected 
light unit and remove the black protective cover. 
You  can  then  insert  the  round  dovetail  bracket  on  the  head  into  the  round  dovetail 
bracket on the reflected light unit and fix it with the fixing screw. When doing this, you 
should always make sure that you do not touch the lenses with your bare fingers and 
that no dust enters the apertures. 

4.3  Objectives 

The  specimen  stage  must  be  in  its  lowest  position  so  that  the  objectives  can  be 
screwed into the nosepiece. You can then screw the objectives into the nosepiece so 
that when you turn the nosepiece in a clockwise direction, the objective with the next 
strongest  magnification  appears.  You  must  make  sure  that  you  do  not  touch  the 
lenses  with  your  bare  fingers  and  that  no  dust  enters  the  apertures.  For  objectives 
which  ar

e  marked  “OIL”,  you  must  use  an  immersion  oil  with  the  lowest  level  of 

inherent fluorescence. 

Summary of Contents for OKN 175

Page 1: ...Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Metallurgical microscope KERN OKN 1 OKO 1 OKN 175 OKN 177...

Page 2: ......

Page 3: ...interpupillary distance 16 5 4 Dioptre adjustment 16 5 5 Adjusting the magnification 17 5 6 Using eye cups 18 5 7 Using oil immersion objectives 19 5 8 Adjusting the Koehler illumination for transmit...

Page 4: ...s an additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The in...

Page 5: ...which are too high or too low vibrations dust or a high level of humidity The ideal temperature range is between 0 and 40 C and a relative humidity of 85 should not be exceeded The device should alway...

Page 6: ...as these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However org...

Page 7: ...OKN 1_OKO 1 BA e 1510 6 2 Nomenclature...

Page 8: ...7 OKN 1_OKO 1 BA e 1510 Rear view...

Page 9: ...OKN 1_OKO 1 BA e 1510 8 Reflected light unit...

Page 10: ...OKO 176 490x200x460 mm Product dimension OKN 177 OKO 178 550x200x460 mm Packing dimensions OKN 175 OKO 176 750x360x450 mm Packing dimensions OKN 177 OKO 178 990x360x450 mm Net weight OKN 175 OKO 176 1...

Page 11: ...OKN 1_OKO 1 BA e 1510 10...

Page 12: ...11 OKN 1_OKO 1 BA e 1510 4 Assembly...

Page 13: ...unit The plug of the connection cable attached to the rear of the light unit has to be installed to the socket on the rear of the power supply unit Please do also use the screw lock on the plug 4 2 M...

Page 14: ...an be removed at any time In order to use suitable filters without this holder ring these can simply be placed in the ring bracket of the field lens 4 6 Condenser OKO 1 We recommend that you use the c...

Page 15: ...now switch on the lighting using the main switch The next step is to place an object holder with the sample on the angle table To do this the cover glass if used must be facing upwards For the models...

Page 16: ...As soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the lef...

Page 17: ...pillary distance is set correctly 5 4 Dioptre adjustment The eye strengths of each eye of the microscope user can often be slightly different which in daily life has no consequences But when using a m...

Page 18: ...al objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam...

Page 19: ...glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would the...

Page 20: ...slide 2 Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the ob...

Page 21: ...hragm Field diaphragm When adjusting the Koehler lighting for the first time you must first select the lowest possible objective magnification so that you can carry out the following steps 1 Use the c...

Page 22: ...the condenser up so high that it collides with the object holder Therefore care is needed when doing this 4 Use the centring screws of the condenser holder to bring the image from the field diaphragm...

Page 23: ...y 2 3 of the pupil diameter If the eyepiece should be removed for checking then please make sure that no dirt or dust falls into the tube 7 It is possible to alter the brightness of the bulb using the...

Page 24: ...2 as well If necessary one of the two apertures has to be snapped into the beam path Polarising unit Analyser Polariser In order to bring the analyser into the beam path you must attach the analyser...

Page 25: ...bjective with the lowest magnification for the beam path 4 Remove all the filter slides from the beam path 5 If there is one remove the object slide from the stage 6 Remove one of the eyepieces from t...

Page 26: ...d The bulb holder is on the underside of the device It can be opened by undoing the screws on the holder see illustration We recommend that here you should also test again to check that heat is no lon...

Page 27: ...t OKO 1 The fuse housing is on the rear of the microscope below the mains power supply socket With the device switched off and unplugged you can pull out the housing When doing this it is helpful to u...

Page 28: ...elow the nosepiece see illustration on page 10 where you can simply insert the analyser see left illustration The polariser on the other hand is only placed on the ring holder of the field lens see ri...

Page 29: ...ide of the microscope head must be pulled out The light from the microscope lighting is deflected so that it is completely in the beam path for the camera which causes a dark field of view in the eyep...

Page 30: ...contrast unit This consists of a simple PH condenser a PH objective with a specific magnification 10x 20x 40x or 100x a PH slider which is adapted to the lens being used a centring telescope and two...

Page 31: ...nces of the user using a green filter can produce a more effective and pleasant image To achieve this the green filter must be screwed onto the underside of the PH condenser 2 Comprehensive phase cont...

Page 32: ...and a black objective ring The black one is central and the white one possible is pushed to one side see left illustration g Now press the turnscrews fitted on the side of the condenser towards the ce...

Page 33: ...ctly centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is betw...

Page 34: ...es Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path T...

Page 35: ...dly materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in for...

Reviews: