background image

31 

 

OKN-1_OKO-1-BA-e-1510 

The adjustment for this is carried out as follows. 

 

a.  First adjust the microscope using bright field mode. 

 

b. 

Move the hub to the required PH position (e.g. “10x”). 

 

c.  Open the aperture diaphragm on the condenser to its maximum (slider to the 

left). 

 

d.  Bring the relevant PH objective (e.g. 10x) into the beam path. 

 

e.  Place the centring telescope onto one of the two tube connectors, in place of a 

standard eyepiece. 

 

f.  Undo the fixing  screw  of  the  centring  telescope and  pull out  (move)  the front 

part  of  the  telescope,  so  that  you  can  focus  both  phase  rings  in  the  field  of 
view.  You can also use the focus wheel of  the condenser holder when doing 
this. Then retighten the screws. 

 
In the field of view you will now see the image of a white (condenser) and a black 
(objective) ring. The black one is central and the white one possible is pushed to 
one side 

(see left illustration)

.  

 

 

 
 
g.  Now  press  the  turnscrews  fitted  on  the  side  of  the  condenser  towards  the 

centre until they grip the screws. Then by turning the screws and at the same 
time watching the rings through the centring telescope, move the white ring to 
the centre 

(see central illustration)

 

h.  As  soon  as  both  the  rings  overlap  each  other 

(see  right  illustration)

  the 

adjustment is successfully completed and can also be carried out for the pairs 
of rings of the other magnifications. 

 

i.  After the adjustment, the centring telescope must be replaced by the standard 

eyepiece,  so  that  you  are  able  to  observe  the  object  on  the  angle  table  in 
phase contrast mode. 

 
 
Depending  on  the  preferences  of  the  user,  using  a  green  filter  can  produce  a 
more  effective  and  pleasant  image.  To  achieve  this,  the  green  filter  must  be 
screwed onto the underside of the PH condenser. 

 

 

Summary of Contents for OKN 175

Page 1: ...Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Metallurgical microscope KERN OKN 1 OKO 1 OKN 175 OKN 177...

Page 2: ......

Page 3: ...interpupillary distance 16 5 4 Dioptre adjustment 16 5 5 Adjusting the magnification 17 5 6 Using eye cups 18 5 7 Using oil immersion objectives 19 5 8 Adjusting the Koehler illumination for transmit...

Page 4: ...s an additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The in...

Page 5: ...which are too high or too low vibrations dust or a high level of humidity The ideal temperature range is between 0 and 40 C and a relative humidity of 85 should not be exceeded The device should alway...

Page 6: ...as these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However org...

Page 7: ...OKN 1_OKO 1 BA e 1510 6 2 Nomenclature...

Page 8: ...7 OKN 1_OKO 1 BA e 1510 Rear view...

Page 9: ...OKN 1_OKO 1 BA e 1510 8 Reflected light unit...

Page 10: ...OKO 176 490x200x460 mm Product dimension OKN 177 OKO 178 550x200x460 mm Packing dimensions OKN 175 OKO 176 750x360x450 mm Packing dimensions OKN 177 OKO 178 990x360x450 mm Net weight OKN 175 OKO 176 1...

Page 11: ...OKN 1_OKO 1 BA e 1510 10...

Page 12: ...11 OKN 1_OKO 1 BA e 1510 4 Assembly...

Page 13: ...unit The plug of the connection cable attached to the rear of the light unit has to be installed to the socket on the rear of the power supply unit Please do also use the screw lock on the plug 4 2 M...

Page 14: ...an be removed at any time In order to use suitable filters without this holder ring these can simply be placed in the ring bracket of the field lens 4 6 Condenser OKO 1 We recommend that you use the c...

Page 15: ...now switch on the lighting using the main switch The next step is to place an object holder with the sample on the angle table To do this the cover glass if used must be facing upwards For the models...

Page 16: ...As soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the lef...

Page 17: ...pillary distance is set correctly 5 4 Dioptre adjustment The eye strengths of each eye of the microscope user can often be slightly different which in daily life has no consequences But when using a m...

Page 18: ...al objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam...

Page 19: ...glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would the...

Page 20: ...slide 2 Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the ob...

Page 21: ...hragm Field diaphragm When adjusting the Koehler lighting for the first time you must first select the lowest possible objective magnification so that you can carry out the following steps 1 Use the c...

Page 22: ...the condenser up so high that it collides with the object holder Therefore care is needed when doing this 4 Use the centring screws of the condenser holder to bring the image from the field diaphragm...

Page 23: ...y 2 3 of the pupil diameter If the eyepiece should be removed for checking then please make sure that no dirt or dust falls into the tube 7 It is possible to alter the brightness of the bulb using the...

Page 24: ...2 as well If necessary one of the two apertures has to be snapped into the beam path Polarising unit Analyser Polariser In order to bring the analyser into the beam path you must attach the analyser...

Page 25: ...bjective with the lowest magnification for the beam path 4 Remove all the filter slides from the beam path 5 If there is one remove the object slide from the stage 6 Remove one of the eyepieces from t...

Page 26: ...d The bulb holder is on the underside of the device It can be opened by undoing the screws on the holder see illustration We recommend that here you should also test again to check that heat is no lon...

Page 27: ...t OKO 1 The fuse housing is on the rear of the microscope below the mains power supply socket With the device switched off and unplugged you can pull out the housing When doing this it is helpful to u...

Page 28: ...elow the nosepiece see illustration on page 10 where you can simply insert the analyser see left illustration The polariser on the other hand is only placed on the ring holder of the field lens see ri...

Page 29: ...ide of the microscope head must be pulled out The light from the microscope lighting is deflected so that it is completely in the beam path for the camera which causes a dark field of view in the eyep...

Page 30: ...contrast unit This consists of a simple PH condenser a PH objective with a specific magnification 10x 20x 40x or 100x a PH slider which is adapted to the lens being used a centring telescope and two...

Page 31: ...nces of the user using a green filter can produce a more effective and pleasant image To achieve this the green filter must be screwed onto the underside of the PH condenser 2 Comprehensive phase cont...

Page 32: ...and a black objective ring The black one is central and the white one possible is pushed to one side see left illustration g Now press the turnscrews fitted on the side of the condenser towards the ce...

Page 33: ...ctly centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is betw...

Page 34: ...es Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path T...

Page 35: ...dly materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in for...

Reviews: