B-18
ATE-DMG 020613
B.77 STATus:OPEReration:ENABle?
QUERY
STAT:OPER:ENAB?
Syntax
:
Short Form: STAT:OPER:ENAB?
Long Form: STATus:OPERation:ENABle?
Return Value: <int_value> 0 to 1313 (1 + 32 + 256 + 1024).
Description:
Reads Operation Enable Register (see Table B-2).
Returns value of Operation Enable Register
bits. Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false).
(See example, Figure B-7.)
B.78 STATus:OPERation[:EVENt] QUERY
STAT:OPER?
Syntax
:
Short Form: STAT:OPER[:EVEN]?
Long Form: STATus:OPERation[:EVENt]?
Return Value: <int_value> 0 to 1313 (1 + 32 + 256 + 1024).
Description:
Indicates changes in conditions monitored by Operational Event Register.
Returns the value of
the Operation Event register. The Operation Event register is a read-only register which holds
(latches) all events that occur. Reading the Operation Event register clears it. (See example, Figure B-
7.)
B.79 STATus:PRESet COMMAND
STAT:PRES
Syntax
:
Short Form: STAT:PRES
Long Form: STATus:PRESet
Description:
Disables reporting of all status events.
This command sets all bits of the Operation Condition
(Table B-2) and Questionable Condition Registers to 0, preventing all status events from being
reported. (See example, Figure B-7.)
FIGURE B-7. USING STATUS COMMANDS AND QUERIES
NOTE:The power supply is assumed to be operating in cV (constant voltage) mode.
STAT:OPER:ENAB 1056Mask enabled for CC, WTG and bits.
STAT:OPER:ENAB? Returns 1056 (32 + 1024) (CC, WTG bits set).
STAT:QUES:ENAB 3 Mask enabled for OV and OC bits (1 + 2).
STAT:QUES:ENAB? Returns 3 (1 + 2) indicating OV and OC bits are enabled.
STAT:PRES
Operation Condition and Questionable Condition registers are
reset.
INIT:CONT ON
Continuous triggers enabled.
STAT:OPER:COND? Power supply returns 288 (256 + 32) to indicate that power
supply is constant voltage mode and Wait For Trigger is true.
STAT:OPER?
Returns 1057, e.g., indicating that since the last reading of the
Operation Event Register the power supply has entered
Constant Current mode, the Wait Trigger was set.
STAT:OPER?
Returns 0 indicating no changes since previous reading of the
Operation Event register.
STAT:QUES?
Returns 0 (no questionable conditions occurred since previous
reading
--- (SHORT OUTPUT) OVERCURRENT CONDITION OCCURS
STAT:QUES?
Returns 2 (overcurrent protection tripped since the last
STAT:QUES? query).
STAT:QUES:COND? Returns 2, (Power supply still in overcurrent protection state).
STAT:QUES?
Returns 0, (Register cleared by previous STAT:QUES?).
STAT:QUES:COND? Returns 2, (Power supply still in overcurrent protection state).
— (REMOVE SHORT FROM OUTPUT)
CURR:PROT:CLE
Overcurrent condition cleared.
STAT:QUES:COND? Returns 0, (register cleared by CURR:PROT:CLE).
SYST:ERR?
Power supply returns 0,“No error” message.
Summary of Contents for ATE 100-10DMG
Page 2: ......
Page 6: ......
Page 14: ......
Page 15: ...ATE DMG020613 ix FIGURE 1 1 ATE DMG SERIES PROGRAMMABLE POWER SUPPLY...
Page 16: ...ix Blank x ATE DMG020613...
Page 24: ...1 8 ATE DMG OPER 020613 FIGURE 1 2 MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Page 26: ...2 2 ATE DMG 020613 FIGURE 2 1 ATE DMG FULL RACK SERIES FRONT PANEL...
Page 34: ......
Page 74: ......
Page 98: ......