ATE-DMG 020613
3-11
output current set to a minimum value (1-2% of I
O
max). Power on defaults also include setting
maximum values for overcurrent and overvoltage protection indicated in
Table 1-4.
FIGURE 3-6. LCD POWER ON DEFAULTS
3.4.3
SETTING LOCAL MODE
When the power supply is turned on, it is automatically set to Local mode. If remote commands
are accepted over the GPIB bus, the power supply will automatically go into Remote mode (
Rem
at the upper left of the LCD). Pressing the
LOCAL
key will restore Local mode. When in Remote
mode, all keys except
LOCAL
are disabled.
The ATE-DMG incorporates a “keypad lockout” command which allows the
LOCAL
key to be
disabled during remote operation, preventing inadvertent setting of the power supply to Local
mode. If the keypad is locked, it must be unlocked either by a remote command (see Appendix
B, PAR. B.87), or cycling the power supply off then on.
3.4.4
ADJUSTING LCD CONTRAST
With the power supply in command entry status , press
MENU
key. Press
"
or
ç
key to
increase or decrease contrast for optimum viewing. The contrast can also be set directly by
entering a number from
1
to
9
. Press
ENTER
or
CLEAR
to exit menu.
3.4.5
ENABLING/DISABLING AUDIBLE BEEPS
With the power supply in command entry status , press
MENU
key three times. The top line
indicates if the speaker is on or off (the factory default setting is speaker on). Enter
1
to enable
the speaker, or
0
to disable the speaker. Press
ENTER
or
CLEAR
to exit menu.
3.4.6
ENABLING/DISABLING DC OUTPUT POWER
When the power supply is turned on, the output is automatically enabled, and the bottom line of
the LCD gives voltage and current measurements at the output. To disable the output, press the
red
OUTPUT ON/OFF
key; the lower line of the LCD reads:
Output OFF
. When the output is
disabled, voltage and current are programmed to zero; the power supply remains in Command
Entry status. Pressing the
OUTPUT ON/OFF
key again enables the output: the previously pro-
grammed values of voltage and current are restored and the LCD displays the mode (CV or CC)
and actual values of output voltage and current.
3.4.7
RESET
RESET
overrides all other local commands to reset the power supply to the power on defaults:
output enabled, voltage mode, output voltage set to zero, output current set to a minimum value
Loc
CV
0.000A
0.000V
NOTE:
indicates blinking colon (:), Command Entry status
indicates blinking equal sign (=), Data Entry status
Summary of Contents for ATE 100-10DMG
Page 2: ......
Page 6: ......
Page 14: ......
Page 15: ...ATE DMG020613 ix FIGURE 1 1 ATE DMG SERIES PROGRAMMABLE POWER SUPPLY...
Page 16: ...ix Blank x ATE DMG020613...
Page 24: ...1 8 ATE DMG OPER 020613 FIGURE 1 2 MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Page 26: ...2 2 ATE DMG 020613 FIGURE 2 1 ATE DMG FULL RACK SERIES FRONT PANEL...
Page 34: ......
Page 74: ......
Page 98: ......