ADJUSTMENT
Measurement
Test-
equipment
Unit
Terminal
Adjustment
Unit
Parts
Method
Specifications/Remarks
Condition
Item
TK-782
Receiver Section
1. Discriminator 1) Set test mode
SSG
Rear
ANT
TX-RX
L6
AF output voltage
CH : CH1 - Sig1
panel
(A/2)
maximum.
SSG output : –53dBm
AF VTVM
ACC
AF : 1.4V/4
Ω
Oscilloscope
(EXT.SP)
2. Sensitivity
1) Set test mode
SSG
ANT
Adjust for
Select “SENS” in tuning mode.
maximum SINAD.
“L SENS”
AF VTVM
ACC
SSG freq’ : 217.050MHz
Distortion
(EXT.SP)
SSG output : –118dBm/0.31
µ
V
meter
SSG MOD : 1.5kHz
Oscilloscope
AF output : 1V/4
Ω
AG
2) “C SENS”
SSG freq’ : 226.050MHz
3) “H SENS”
SSG freq’ : 234.950MHz
3. Squelch
1) Set test mode
Squelch must be
Select “nL SQL” in tuning mode.
closed once.
SSG freq’ : 217.050MHz
Then adjust for
SSG output : Sensitivity
squelch open point.
value of 12dB SINAD.
SSG MOD : 1.5kHz (Narrow)
2) “nC SQL”
SSG freq’ : 226.050MHz
3) “nH SQL”
SSG freq’ : 234.950MHz
4) “wL SQL”, “wC SQL”, “wH SQL”
SSG freq’ : Same as
narrow adjustment.
SSG output : Sensitivity
value of 12dB SINAD.
SSG MOD : 3.0kHz (Wide)
4. RSSI (Low)
1) Set test mode
Writing values only.
Select “nL LRSI” in tuning mode.
SSG freq’ : 217.050MHz
SSG output : Sensitivity
value of 12dB SINAD.
SSG MOD : 1.5kHz (Narrow)
2) “nC LRSI”
SSG freq’ : 226.050MHz
66