User Manual 

KS-TB-100 | Portable Turbidity 

Meter 

 

Kensington Scientific 

22a Cavan Rd, Dry Creek 
South Australia, 5094 
[email protected] 
www.kensingtonscientific.com.au

 

OVERVIEW 
 

Below are the steps you need to follow to begin using your Kensington KS-TB-100 portable turbidity meter. This 
manual provides a step-by-step guide to help you operate the meter. 
 

UNPACKING METER CONTENTS 

 
Before unpacking the meter and its contents, ensure that the working conditions meet the following conditions: 
 
1. Relative humidity is less than 80%. 
2. Ambient temperature is greater than °C and less than 60°C 
3. No potential electromagnetic interference. 
 

INCLUDED ACCESSORIES 

 
The KS-TB-100 portable turbidity meter includes; 
 
1. Glass sample vials  
2. Calibration standards 0, 10, 200 and 1000 NTU  
3.  Lint-free cloth  
4. 9V battery 
 

OPERATING PRINCIPLES 
 

The KS-TB-100 portable turbidity meter operates on the nephelometric principle of turbidity measurement and is 
designed to meet the criteria specified in ISO 7027.  
 
The following figure shows basic optical system of the meter. It includes a light source and a detector to monitor the 
light scattered at 90° with respect to the incident beam. 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for KS-TB-100

Page 1: ...lowing conditions 1 Relative humidity is less than 80 2 Ambient temperature is greater than C and less than 60 C 3 No potential electromagnetic interference INCLUDED ACCESSORIES The KS TB 100 portable...

Page 2: ...gtonscientific com au OVERVIEW The KS TB 100 measures turbidity from 0 to 1100 NTU FNU or 0 to 275 EBC or 0 to 9999 ASBC The setup menu contains the 7 optional parameters Data can be stored and transf...

Page 3: ...bration icon Indicates meter is calibrating KEYPAD Key Description 1 Switching the turbidity meter On or Off 2 Press the key to perform a measurement 3 In the calibration or setting mode exits current...

Page 4: ...tric Turbidity Unit FNU Formazin Nephelometric Unit EBC European Brewery Convention ASBC American Society if Brewing Chemistries RESO Sets the resolution for measurement 0 1 Resolution at 0 1 0 01 Def...

Page 5: ...m until the meter returns to measurement mode Setting is completed SETTING THE TURBIDITY UNIT 1 Press and hold the CAL key for three seconds to enter setup mode 2 Press or key until the display show U...

Page 6: ...meter should show 0 1 rESO 4 Press or key to select the desired resolution 5 Press ENTER to confirm The meter should return to measurement mode AUTO POWER OFF 1 Press and hold the CAL key to enter se...

Page 7: ...e is selected 5 Select the desired measurement mode press ENTER key to confirm The meter returns to measurement mode CALIBRATION GUIDELINES FOR MEASUREMENT AND CALIBRATION 1 Do not hold the meter in t...

Page 8: ...ent light interference do not use the meter in bright sunlight Always close the light shield lid during measurement and calibration 8 Avoid spillage of water in the sample chamber It may cause damage...

Page 9: ...ng that the meter will perform the 2 points calibration 2 If necessary press or to select the number of calibration points 2 to 5 points 3 Press ENTER key to confirm the display shows 0 02 NTU CAL1 an...

Page 10: ...stops flashing 8 Press or when needed to modify the calibration value 9 Press ENTER key the vial icon begins flashing 10 Wait for the measured value to stabilize the vial icon will disappear on the di...

Page 11: ...MENT FOR LOW TURBIDITY SAMPLES 1 Rinse the vial with approximately 10 ml of the sample capping the vial with the screw cap and gently inverting it several times Discard the used sample and repeat the...

Page 12: ...ral times to mix the sample 4 Wipe the vial with the lint free cloth to remove waterdrops and fingerprints Ensure that the outside of the vial is dry and clean 5 Place the vial in the sample chamber A...