THE CALIBRATION PROCESS
1. Press CAL key, the display shows
“2/CAL” indicating that the meter will perform the 2 points calibration
2. If necessary, press ◄ or ► to select the number of calibration points (2 to 5 points).
3. Press ENTER key to confirm, the display shows “0.02 NTU/CAL1” and continuously flashing.
4. Press ◄ or ► to select desired calibration point (e.g., 10.00 NTU).
5. Insert the corresponding standard into the sample chamber, ensure that the mark on the vial cap always aligns with
the arrow on the meter.
6. Close the light shield lid.