7. Press ENTER key, the displayed value stops flashing.
8. Press ◄ or ► when needed to modify the calibration value.
9. Press ENTER key, the “vial” icon begins flashing.
10. Wait for the measured value to stabilize, the “vial” icon will disappear on the display.
11. Press ENTER key to confirm.
12. The display shows “0.02 NTU/CAL2” and continuously flashing.
13. Repeat steps 4 to 11 above until the display shows “End”, the meter returns to measurement mode. Calibration is
completed.