![background image](http://html.mh-extra.com/html/kensington/ks-sal-20/ks-sal-20_user-manual_664415001.webp)
User Manual
KS-SAL-20 |Pocket
Salinity Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
[email protected]
www.kensingtonscientific.com.au
OVERVIEW
Below are the steps you need to follow to begin using your Kensington KS-SAL-20 pocket salinity tester. This manual
will cover how insert the batteries, using the function keys, making measurements, calibration and sensor replacement.
PACKAGE CONTENTS
1. Conductivity calibration solutions (1413 µS/cm, 12.88m µS/cm, 111.8 S/cm,)
2. AAA batteries
3. Carrying case
INSERTING THE BATTERIES
1. Your Kensington Pocket KS-SAL-20 pocket salinity tester uses two AAA batteries.
2. Remove the button cap found on the bottom of the tester as shown in the above image.
3. Place the battery into its designated slot. Take note of the proper position of the battery to ensure the polarities
placed correctly.
4. Replace the battery cap.
KEYPAD INDEX
Key
Function
1. Serves as the power button of the tester
2. Initiates Data Hold. Press once to hold the current reading, press again to resume making
measurements.
3. Exit button during calibration
1. Press once to conduct calibration of the instrument
2. Press and hold to enter setup menu
3. In calibration mode, press to set calibration values.
4. In setup mode, press to select default option