background image

 

 

 

User Manual 

KS-PH-220-BASIC | Portable 

pH/mV Meter 

 

Kensington Scientific 

22a Cavan Rd, Dry Creek 
South Australia, 5094 
[email protected] 
www.kensingtonscientific.com.au

 

MEMORY RECALL 

 
1. Press the VIEW key in measurement mode. The display should show the ID number, data and time of the stored  
    data. 
   
 
 
 
 
 
 
 
 
 
 

2. Press the    key to scroll down on the stored data, showing either the pH or mV reading and temperature. 
 
 
 
 
 
 
 
 
 
 
 
 
3. To exit, press MEAS key to return to measurement mode. 
 

ELECTRODE CARE AND MAINTENANCE 

 
1. Clean as necessary depending on the extent and condition of use. Store the electrode in electrode storage solution 
    after cleaning or after use. 
2. When used in samples with salt deposits: Wash the electrode using tap water and then rinse using distilled water. 
3. When used in samples with oil and grease: Was the bulb gently in water with detergent. Using alcohol is also  
    effective. 
4. Clogged reference junction: Heat a diluted KCl solution to 60 to 80 

°C and place the senor into the heated solution  

    for 10 minutes. Allow the electrode to cool in  some unheated KCl solution. 
5. For solutions with protein deposits: Prepare a 1% pepsin solution in 0.1M of HCL. Place the electrode in the solution  
    for 10 minutes. Rinse the sensor with distilled water 
 

PREPARATION OF THE ELECTRODE STORAGE SOLUTION 

 
1. Dissolve 22.365 grams of potassium chloride reagent (KCL) in the 100mL distilled water. 
2. Pour the 50mL of KCL solution into the beaker. 
3. Pour the 50mL of pH4.01 standard buffer solution into the same beaker. 
4. Stir the solution until the solution has thoroughly mixed. Preparation is completed. 

 
 
 

Summary of Contents for KS-PH-220-BASIC

Page 1: ...Temperature Probe 3 pH Buffer Solutions pH4 01 7 00 10 01 4 9V Battery 5 Carrying Case DISPLAY The KS PH 220 BASIC is built with a large and clear LCD display that shows measured values mode indicator...

Page 2: ...e key to increase the value by 1 3 In memory mode press to search the previous saved measurements 1 In the measurement mode press this key to recall stored data in the memory 2 In the temperature sett...

Page 3: ...set the time and press ENTER to confirm The settings are now saved and the meter will then return to measurement mode TEMPERATURE COMPENSATION Temperature compensation is needed to make accurate measu...

Page 4: ...ation point the pH 7 00 buffer solution is used The solution used for the second calibration point depends on the sample to be measured For acidic samples the pH 4 01 buffer solution should be used Fo...

Page 5: ...completed The display will then show indicating the start of the second calibration point 7 Begin the second calibration point by rinsing the pH electrode with distilled water 8 Dip the electrode into...

Page 6: ...ove impurities 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the measured value to stabilize and record the reading on the display HOLD FUNCTION The data hold function...

Page 7: ...h the electrode using tap water and then rinse using distilled water 3 When used in samples with oil and grease Was the bulb gently in water with detergent Using alcohol is also effective 4 Clogged re...

Page 8: ...pH Resolution 0 01 pH Calibration points 2 points Calibration solutions USA pH4 01 7 00 10 01 Temperature compensation 0 100 C 32 212 F Manual or Automatic mV Range 1999 1999 mV Accuracy 1 mV Resoluti...

Reviews: