KEYPAD 
 

The KS-PH-220-BASIC has six (6) multi-function keys. Below are their names and functions 
 

Key 

Description 

 

1. Press this button to activate HOLD function 
2. Freezes the measured value on the display, press the key again to resume measuring. 
3. In the calibration or setting mode, exits current mode and returns to measurement. 

 

Press to enter SETUP mode 

 

Press to activate temperature setting mode 

 

1. In the measurement modem press this key to store current measured value. 
2. In the temperature setting mode, pressing the key increases the value by 0.1. Press and hold the  
    key to increase the value by 1. 
3. In memory mode, press   to search the previous saved measurements 

 

1. In the measurement mode, press this key to recall stored data in the memory 
2. In the temperature setting mode, press   to search the next stored data 

 

1. Toggles between pH and mV measurement modes 
2. In the setup mode, press this key to confirm option or input values. 

 

CONNECTING THE pH ELECTRODE 
 
 
 

 
 
 
 
 
 
 
 
 
1. Take out the pH electrode from the carrying case. 
2. Insert the connector of the electrode into the BNC connector socket on the meter. Rotate and push the connector  
    until it locks. 
3. Do not pull on the sensor cord once the electrode is connected to the meter. 
 

USER SETTINGS 
 
SELECTING THE TEMPERATURE UNIT 

 
1. Press SETUP key to enter the setup menu 

2. Press   or   

key to select the ‘Temperature Units’ option 

3. Press ENTER key to confirm. The ‘C’ indicator should begin flashing. 
4. Press   or   keys to select the desired temperature units (°C or °F) 
5. Press ENTER to confirm. 
 
 
 
 
 

Summary of Contents for KS-PH-220-BASIC

Page 1: ...Temperature Probe 3 pH Buffer Solutions pH4 01 7 00 10 01 4 9V Battery 5 Carrying Case DISPLAY The KS PH 220 BASIC is built with a large and clear LCD display that shows measured values mode indicator...

Page 2: ...e key to increase the value by 1 3 In memory mode press to search the previous saved measurements 1 In the measurement mode press this key to recall stored data in the memory 2 In the temperature sett...

Page 3: ...set the time and press ENTER to confirm The settings are now saved and the meter will then return to measurement mode TEMPERATURE COMPENSATION Temperature compensation is needed to make accurate measu...

Page 4: ...ation point the pH 7 00 buffer solution is used The solution used for the second calibration point depends on the sample to be measured For acidic samples the pH 4 01 buffer solution should be used Fo...

Page 5: ...completed The display will then show indicating the start of the second calibration point 7 Begin the second calibration point by rinsing the pH electrode with distilled water 8 Dip the electrode into...

Page 6: ...ove impurities 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the measured value to stabilize and record the reading on the display HOLD FUNCTION The data hold function...

Page 7: ...h the electrode using tap water and then rinse using distilled water 3 When used in samples with oil and grease Was the bulb gently in water with detergent Using alcohol is also effective 4 Clogged re...

Page 8: ...pH Resolution 0 01 pH Calibration points 2 points Calibration solutions USA pH4 01 7 00 10 01 Temperature compensation 0 100 C 32 212 F Manual or Automatic mV Range 1999 1999 mV Accuracy 1 mV Resoluti...