SETTING THE REAL TIME CLOCK 
 

1. Press SETUP key to enter the setup menu. 

2. Press   or   

key to select ‘Date and Time’ option. 

3. Press ENTER key to confirm, the date digit will then start to flash 
 
 
 
 
 
 
 
 
 
 
 

4. Press   or   key to set the data and press ENTER to confirm. The hour digit will then start to flash 

5. Press   or   key to set the time and press ENTER to confirm. The settings are now saved and the meter will then  
    return to measurement mode. 
 

TEMPERATURE COMPENSATION 

 
Temperature compensation is needed to make accurate measurement results. You need to choose between automatic 
and manual temperature compensation before measurement or calibration. 
 

AUTOMATIC TEMPERATURE COMPENSATION 

 
1. Automatic temperature compensation will readily activate when the temperature probe is inserted into the 

meter’s  

    phone jack. 
2. Insert the connector of the temperature probe into the meter’s phone jack 
 
 
 
 
 
 
 
 
 
 
 
 
 
3. The ‘ATC’ indicator should immediately show on the display, indicating that automatic temperature compensation is  
    now enabled.  
 
 
 
 
 
 

Summary of Contents for KS-PH-220-BASIC

Page 1: ...Temperature Probe 3 pH Buffer Solutions pH4 01 7 00 10 01 4 9V Battery 5 Carrying Case DISPLAY The KS PH 220 BASIC is built with a large and clear LCD display that shows measured values mode indicator...

Page 2: ...e key to increase the value by 1 3 In memory mode press to search the previous saved measurements 1 In the measurement mode press this key to recall stored data in the memory 2 In the temperature sett...

Page 3: ...set the time and press ENTER to confirm The settings are now saved and the meter will then return to measurement mode TEMPERATURE COMPENSATION Temperature compensation is needed to make accurate measu...

Page 4: ...ation point the pH 7 00 buffer solution is used The solution used for the second calibration point depends on the sample to be measured For acidic samples the pH 4 01 buffer solution should be used Fo...

Page 5: ...completed The display will then show indicating the start of the second calibration point 7 Begin the second calibration point by rinsing the pH electrode with distilled water 8 Dip the electrode into...

Page 6: ...ove impurities 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the measured value to stabilize and record the reading on the display HOLD FUNCTION The data hold function...

Page 7: ...h the electrode using tap water and then rinse using distilled water 3 When used in samples with oil and grease Was the bulb gently in water with detergent Using alcohol is also effective 4 Clogged re...

Page 8: ...pH Resolution 0 01 pH Calibration points 2 points Calibration solutions USA pH4 01 7 00 10 01 Temperature compensation 0 100 C 32 212 F Manual or Automatic mV Range 1999 1999 mV Accuracy 1 mV Resoluti...