31
[Removal of globe and sensor cap B]
Request
Correct measurements may not be possible if the sensor element is touched. Be sure to install
the sensor without touching it.
The sensor cap may be broken if the globe is turned too hard when installing it. Be sure to
Install the sensor cap at the position where the screw stops and the connections between
the sensor cap and the globe are aligned as a guide. (Refer to above figures)
Make sure not to pull the temperature/humidity sensor unit out during maintenance.
Screw
1.
Hold the globe and the main unit,
and turn the globe counterclockwise
to remove it.
Be careful not to drop the globe when
removing it.
2.
Slide the sensor cap B on the back
of the main unit slightly upward and
remove it.
3.
Install the sensor cap B and the
globe in the reverse order of removal
after drying the temperature/humidity
sensor unit (especially the sensor
element) with a blower.
Make sure that the globe packing fit
in the recess of the sensor cap.
Turn the globe to the position where
it stops and install it.
※
Do not remove the screws.
Sensor element
(opposite side)
Temperature/humidity
sensor unit
Globe
Sensor cap B
Globe packing
Position to align
as a guide
Back side
Serial No.