Quick Start Guide
Integra Series 15
Offset compensated (OCOMP) ohms measurements
For more accurate low resistance measurements, all Integra systems provide the
offset compensation mode to eliminate errors from the thermoelectric EMF
effects. During the measurement cycle, the built-in ohms current source is turned
off, and then turned on again, and the resulting EMF error is automatically sub-
tracted. This technique is typically used when measuring values less than 100
Ω.
Use 4-wire connections, as shown in
B (a).
With the
Ω
4 function selected, OCOMP is enabled buy pressing the SHIFT key
and then the
Ω
4 (OCOMP) key. When enabled, the “OCOMP” annunciator is on.
Valid measurement ranges for OCOMP is 10k
Ω
or less, or use Auto range.
OCOMP is disabled by again pressing SHIFT and then
Ω
4.
SCPI programming commands (volts, amps, and ohms)
DC Volts
SENS:FUNC ‘VOLT:DC’
‘ Select DC Volts function.
SENS:VOLT:DC:RANG:AUTO <b> ‘ Enable or disable auto range: <b> = ON or OFF.
SENS:VOLT:DC:RANG <n>
‘ To set range, specify expected magnitude of input:
‘ <n> = 0 to 1010 (V).
AC Volts
SENS:FUNC ‘VOLT:AC’
‘ Select AC Volts function.
SENS:VOLT:AC:RANG:AUTO <b> ‘ Enable or disable auto range: <b> = ON or OFF.
SENS:VOLT:AC:RANG <n>
‘ To set range, specify expected magnitude of input:
‘ <n> = 0 to 757.5 (V).
DC Amps
SENS:FUNC ‘CURR:DC’
‘ Select DC Amps function.
SENS:CURR:DC:RANG:AUTO <b> ‘ Enable or disable auto range: <b> = ON or OFF.
SENS:CURR:DC:RANG <n>
‘ To set range, specify expected magnitude of input:
‘ <n> = 0 to 3 (A).
AC Amps
SENS:FUNC ‘CURR:AC’
‘ Select AC Amps function.
SENS:CURR:AC:RANG:AUTO <b> ‘ Enable or disable auto range: <b> = ON or OFF.
SENS:CURR:AC:RANG <n>
‘ To set range, specify expected magnitude of input:
‘ <n> = 0 to 3 (A).
2-Wire Ohms
SENS:FUNC ‘RES’
‘ Select 2-Wire Ohms function.
SENS:RES:RANG:AUTO <b>
‘ Enable or disable auto range: <b> = ON or OFF.
SENS:RES:RANG <n>
‘ To set range, specify expected resistance at input:
‘ <n> = 0 to 120e6 (
Ω
).
4-Wire Ohms
SENS:FUNC ‘FRES’
‘ Select 4-Wire Ohms function.
SENS:FRES:RANG:AUTO <b>
‘ Enable or disable auto range: <b> = ON or OFF.
SENS:FRES:OCOM<b>
‘ Enable or disable offset compensated ohms.
SENS:FRES:RANG <n>
‘ To set range, specify expected resistance at input:
‘ <n> = 0 to 120e6 (
Ω
).
To
p
To
p
To
p
Summary of Contents for Integra 27 Series
Page 3: ......
Page 8: ...QuickStartGuide Integra Series...