Connecting Signals
4-13
To prevent this problem, use the following precautions:
●
Avoid direct connections to the AC line.
●
Make sure all connections are tight and sound so that signal wires
will not come loose and short to high voltages.
●
Use isolation amplifiers where necessary.
Connecting a Signal to a Single-Ended Analog Input
Figure 4-12 shows the connections between a signal source and a channel
of a DAS-1800ST/HR Series board configured for single-ended inputs.
For information on single-ended ground connections, refer to “Low-Side
Reference Selection for Single-Ended Inputs” on page 2-6.
Figure 4-12. Wiring a Signal Source to a DAS-1800ST/HR Series Board
Configured for Single-Ended Inputs
Note:
When wiring signals to the analog input channels, you are advised
to wire all unused channels to LL GND or U_CM MD to prevent the
input amplifiers from saturating and ensure the accuracy of your data.
Channel n High
Signal
Source
+
-
LL GND or
U_CM MD
DAS-1800ST/HR
Series Board
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com