A-8
Specifications
Absolute accuracy
±2 LSB @ 25˚C
Temperature drift
Offset: ±5ppm/˚C maximum
Gain: ±5ppm/˚C maximum
Glitch energy
Area: 15nV * seconds
Peak: 15mV
Capacitive load drive
100µF
Settling time
100µs to ±1 LSB of final value (less than 100pF load)
Output current drive
±5mA (25mA short-circuit current)
Power up
DACs power up to 0.0V at reset
Table A-5. Digital I/O Specifications
Attributes
DAS-1800ST/HR Series
Digital output signals
DOSTB, SSHO, and TGOUT:
• V
OH
(min.) = 2.4V @ I
OH
= -3mA
• V
OL
(max.) = 0.5V @ I
OL
= 24mA
Digital output signals
DO 0 to DO 3, GEXT, and MUX 3 to MUX 7:
• V
OH
(min.) = 2.7V @ I
OH
=
−
400µA
• V
OL
(max.) = 0.5V @ I
OL
= 8mA
Digital input signals
DI 0 to DI 3
1
, XPCLK and TGIN
2
:
• V
IH
(min.) = 2.0V; I
IH
(max.) = 20µA
• V
IL
(max.) = 0.8V; I
IL
(max.) =
−
0.2mA
1
Digital inputs DI 0 to DI 3 are pulled up with a 10k
Ω
resistor.
2
On the DAS-1800ST/HR Series basic models, inputs TGIN and XPCLK are not pulled up; on the DA
models, inputs TGIN and XPCLK are pulled up with 10k
Ω
resistors.
Digital output strobe pulse
width
300ns typical; data is latched on the rising edge of DOSTB
Table A-4. Analog Output Specifications for DAS-1802HR-DA (cont.)
Attribute
Specification
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com