Connecting Signals
4-11
Notes:
If you are using a DAS-1802HR or DAS-1802HR-DA board with
an EXP-1800 expansion accessory, it is recommended that you use the
EXP-1800 at a gain of 1 only; using the EXP-1800 at a gain of 50 can
produce less than satisfactory resolution.
If you are using a DAS-1802HR or DAS-1802HR-DA board, it is
recommended that you use only one EXP-1800 expansion accessory;
using more than one EXP-1800 can reduce the performance of the board.
Connecting Signals
This section contains precautionary advice to consider before making I/O
connections. The section also shows some circuits for wiring signal
sources to input channels of DAS-1800ST/HR Series boards.
The circuit diagrams represent a single signal source wired to a single
channel (channel
n
). In reality, you can wire eight separate signal sources
to eight differential inputs or 16 separate signal sources to 16
single-ended inputs.
DAS-1800ST/HR Series boards contain separate grounds for low-level
analog, ±15V power return, and digital signals. An analog ground
(LL GND or U_CM MD, depending on the input configuration) is for
analog signals, a ±15V return is for analog power, and a digital ground
(DGND) is for digital signals and the +5V power-supply return. If using a
differential input configuration, use LL GND for your analog ground. If
using a single-ended input configuration, use LL GND or U_CM MD for
your analog ground; refer to “Low-Side Reference Selection for
Single-Ended Inputs” on page 2-6 for information on choosing between
available analog grounds.
Precautions
If you expect to use a DAS-1801ST or DAS-1801ST-DA board at high
input gain, read the precautionary information in the following section.
Other considerations for I/O connections are offered under “Additional
Precautions” on page 4-12.