Operation
5-15
Ground loops are not normally a problem with instruments
having isolated LO terminals. However, all instruments in
the test setup may not be designed in this manner. When in
doubt, consult the manual for all instrumentation in the test
setup.
Keeping connectors clean
As is the case with any high-resistance device, the integrity
of connectors can be damaged if they are not handled prop-
erly. If connector insulation becomes contaminated, the
insulation resistance will be substantially reduced, affecting
high-impedance measurement paths.
Oils and salts from the skin can contaminate connector
insulators, reducing their resistance. Also, contaminants
present in the air can be deposited on the insulator surface.
To avoid these problems, never touch the connector insulat-
ing material. In addition, the multiplexer-digital I/O card
should be used only in clean, dry environments to avoid
contamination.
If the connector insulators should become contaminated, ei-
ther by inadvertent touching, or from airborne deposits, they
can be cleaned with a cotton swab dipped in clean methanol.
After thoroughly cleaning, they should be allowed to dry for
several hours in a low-humidity environment before use, or
they can be dried more quickly using dry nitrogen.
AC frequency response
The AC frequency response of the Model 7021 is important
in test systems that switch AC signals. Refer to the specifi-
cations at the front of this manual.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7021
Page 86: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 87: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 88: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 89: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 90: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...