Card Connections and Installation
4-10
Single-card system
Figure 4-13 shows how external connections can be made to a
single-card system that uses the multi-pin connector card. This
single-card system is configured as dual 1
×
12 and 1
×
18
multiplex banks.
Instrumentation and DUTs are hard-wired to the Model
7011-MTR male bulkhead connector. This connector has
solder cups that will accept wire size up to #24 AWG. The
test system is connected to the multiplexer using the Model
7011-MTC-2 round cable assembly. This cable mates direct-
ly to both the external bulkhead connector and the Model
7021. Notice that the bulkhead connector is shown mounted
to a fixture to help keep the cabling stable during the test.
Figure 4-13
Single-card system example
7021
Instrument
Instrument
HI and LO to
Bank A
HI and LO to
Bank B
Fixture for
Bulkhead
Connector
7011-MTR
Bulkhead
Connector
7011-MTC-2
Cable Assembly
64 Individual Conductors
DUT Test Fixture
1
12 1
18
Instruments
DUTs
1
12
1
12
1
18
1
18
Simplified Equivalent Circuit
7021
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7021
Page 86: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 87: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 88: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 89: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 90: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...