![background image](http://html2.mh-extra.com/html/keithley/4210-mmpc-s/4210-mmpc-s_quick-start-manual_5096714007.webp)
Model 4210 MMPC-S Quick Start Guide
6
PA-1000 Rev. B / January 2010
Installing the prober cable kits; basic cable setup
Figure 7 shows the basic cable setup for the 2-pin I-V and C-V testing scenarios described in this guide. The 4-pin
I-V and 4-pin C-V testing scenarios require minor setup changes; pulse I-V setup, which is more complex, is
described later in this quick start guide.
Figure 7: Basic 2-pin prober cable setup
Installation guidelines
Figure 8 shows the prober cable assemblies of two Model 4200-MMPC-S kits installed in a Suss Micro Tec
PA300 prober.
Figure 8: Two prober cable assemblies installed in Suss prober
•
Each prober cable assembly is secured to a manipulator by a single mounting screw. Figure 8 shows the
location of the mounting screws. Firmly tighten the cable assemblies to the manipulators.