The guide contains information about the installation and use of this cable kit: 

 

Connecting Model 4200 

instrumentation to the prober bulkhead: Explains how to make connections from 

the Model 4200 to the prober bulkhead. 

 

Installing the prober cable kits

: Explains how to make connections from the prober bulkhead to the prober 

pins.  

 

Usage scenarios:

 Provides the specific prober cable setups for I-V testing (2-wire and 4-wire) and C-V 

testing (2-wire and 4-wire). It also provides the setup for performing pulse I-V measurements using the 
Keithley Model 4200-PIV-A.  

Related documents 

The following documents (in PDF format) are located on the Model 4200 Complete Reference CD:  

 

PA-993:

 This is a one-page packing list for the Model 4210-MMPC-S Multi-Measurement Prober Cable Kit 

for the Suss Micro Tec PA200/300 series prober. A photo shows all the components of the cable kit.  

 

PA-994:

 This is a one-page packing list for the Model 4210-MMPC-C Multi-Measurement Prober Cable Kit 

for the Cascade MicroTech 12000 series prober. A photo shows all the components of the cable kit. 

 

White Paper

 – Labs’ Demands for Greater Measurement Flexibility Require Cabling Systems Capable of 

Accommodating Multiple Measurement Types: This white paper explains the different cabling requirements 
for I-V, C-V, and pulse I-V testing. It also describes an easy-to-use single multi-measurement cabling 
system (Model 4210-MMPC-S for the Suss Micro Tec PA200/300 prober). 

Connecting Model 4200 instrumentation to the prober bulkhead 

WARNING 

To prevent injury or death due to electric shock, remove all power from the 
Model 4200-SCS and shut down the Suss prober before installing the prober 
cable kit.  

The Model 4200 connects to the female triax connectors on the prober bulkhead. Figure 2 shows the triax 
connectors on one of the two connection panels of the prober.  

Figure 2: Prober bulkhead female triax connectors 

 

Summary of Contents for 4210-MMPC-S

Page 1: ...c i a n sw i t ho v e r 1 5 0y e a r so f s p e c i a l i z a t i o n 9 0d a yw a r r a n t y 5d a yr i g h t o f r e t u r no na l l e q u i p me n t 1 3y e a r w a r r a n t i e sf o r n e wa n d p...

Page 2: ...able setup It is also used with the Keithley Model 4200 PIV A package to perform pulse I V measurements This kit has been assembled for use with the Suss Micro Tec PA200 300 prober series NOTE For C V...

Page 3: ...PA200 300 series prober A photo shows all the components of the cable kit PA 994 This is a one page packing list for the Model 4210 MMPC C Multi Measurement Prober Cable Kit for the Cascade MicroTech...

Page 4: ...s two Model 4200 SMU prober connections NOTE Figure 3 shows SMUs that do not use preamps These SMUs use the supplied triax cables that are terminated with a miniature triax connector on one end and a...

Page 5: ...and other components Use the supplied cables and connectors to connect the Model 4200 to the bulkhead of the prober see Figures 5 and 6 Figure 5 shows the Model 4200 connections to the prober and the...

Page 6: ...Model 4210 MMPC S Quick Start Guide PA 1000 Rev B January 2010 5 Figure 5 Model 4200 PIV A instrumentation connections Figure 6 Cables and adapters used for instrumentation connections...

Page 7: ...setup which is more complex is described later in this quick start guide Figure 7 Basic 2 pin prober cable setup Installation guidelines Figure 8 shows the prober cable assemblies of two Model 4200 M...

Page 8: ...od practice to install the second jumper It not only improves the connection but it is a good place to store the extra jumper Figure 9 shows another view of the jumper Figure 9 Front view of the insta...

Page 9: ...m 2 pin I V testing Figure 3 shows the SMU connections to the outside of the bulkhead Figure 10 Test setup for 2 pin I V testing I V testing 4 pin I V setup With four Model 4200 SMUs connected to the...

Page 10: ...pin C V setup NOTE The setup for 2 pin C V testing requires two prober cable kits With a Model 4210 CVU connected to the prober bulkhead use the fundamental setup shown in Figure 12 to perform 2 pin...

Page 11: ...s a factory installed package that includes a scope card pulse generator card software and other components to perform 4 pin pulse I V measurements The test setup for using the Model 4200 PIV A is sho...

Page 12: ...Model 4210 MMPC S Quick Start Guide PA 1000 Rev B January 2010 11 Figure 14 Connections to the gate and source of the DUT Figure 15 Connections to the drain and bulk Figure 16 Jumper connections...

Reviews: