Model 2701 User’s Manual
SCPI Reference Tables
15-9
[SENSe[1]]
:VOLTage[:DC]
Path to configure DC voltage.
:APERture <n> [, <clist>]
Set integration rate in seconds
(3.333333e-5 to 1).
(Note 2)
:APERture? [<clist>]
Query aperture integration rate.
:NPLCycles <n> [, <clist>]
Set integration rate in line cycles (60Hz;
0.002 to 60, 50Hz; 0.002 to 50).
5.0
:NPLCycles? [<clist>]
Query line cycle integration rate.
:RANGe
Path to set measurement range:
[:UPPer] <n> [, <clist>]
Select range (0 to 1010).
1000
[:UPPer]? [<clist>]
Query range.
:AUTO <b> [, <clist>]
Enable or disable auto range.
ON
:AUTO? [<clist>]
Query state of auto range.
:DIGits <n> [, <clist>]
Specify measurement resolution (4 to 7).
7
:DIGits? [<clist>]
Query resolution.
:REFerence <n> [, <clist>]
Specify reference (-1010 to 1010).
0
:STATe <b> [, <clist>]
Enable or disable reference.
OFF
:STATe? [<clist>]
Query state of reference.
:ACQuire [, <clist>]
Use input signal as reference.
:REFerence? [<clist>]
Query reference value.
:AVERage
Path to configure and control filter:
:TCONtrol <name>
Select filter type: (MOVing or REPeat).
(Note 3)
:TCONtrol?
Query filter type.
:WINDow <NRf>
Set filter window in % of range
(0 to 10).
0.1
:WINDow?
Query filter window.
:COUNt <n> [, <clist>]
Specify filter count (1 to 100).
10
:COUNt? [<clist>]
Query filter count.
[:STATe] <b> [, <clist>]
Enable or disable filter.
(Note 4)
[:STATe]? [<clist>]
Query state of digital filter.
:IDIVider <b>
Enable or disable 10M
Ω
input divider.
OFF
:IDIVider?
Query state of input divider.
Table 15-5 (continued)
SENSe command summary
Command
Description
Default
parameter
Ref
SCPI
2701-900-01.book Page 9 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...