4-16
Troubleshooting
No comm link error
A “No Comm Link” error indicates that the front panel processor has stopped communicating
with the main processor, which is located on the motherboard. This error indicates that one of
the main processor ROMs may require re-seating in its socket. ROMs may be reseated as
follows:
1.
Turn off the power, and disconnect the line cord and all other test leads and cables from
the instrument.
2.
Remove the case cover as outlined in
.
3.
Locate the two firmware ROMs, U102 and U103, located on the motherboard. These are
the only ICs installed in sockets. (Refer to the component layout drawing at the end of
for exact locations.)
4.
Carefully push down on each ROM IC to make sure it is properly seated in its socket.
CAUTION
Be careful not to push down excessively, or you might crack the
motherboard.
5.
Connect the line cord, and turn on the power. If the problem persists, additional
troubleshooting will be required.
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......