Remote factory calibration
Follow the steps in this section to perform factory calibration via remote. See
a detailed list and description of SCPI calibration commands.
When sending calibration commands, be sure that the Model 2182 completes each step before
sending the next command. You can do so either by observing the front panel PRECAL or
CALIBRATING message, or by detecting the completion of each step over the bus. Be sure to
include a space character between each command and parameter.
The procedures for factory calibrating the Model 2182 using SCPI commands include:
•
Preparing the Model 2182 for factory calibration
•
Precalibration
•
Front panel short and open calibration
•
DC volts calibration
•
Analog output calibration
•
Programming calibration dates
•
Saving calibration constants
•
Locking out calibration
Preparing the Model 2182 for factory calibration
1.
Connect the Model 2182 to the IEEE-488 bus of the computer using a shielded
IEEE-488 cable, such as the Keithley Model 7007, or connect the unit to a computer
through the RS-232 port using a straight-through 9-pin to 9-pin cable (use a 9-25-pin
adapter if necessary).
2.
Press in and hold the STEP key, then turn on the Model 2182, and allow it to warm up
for 2-1/2 hours before performing calibration.
3.
Select the DCV1 function.
4.
Make sure the primary address of the Model 2182 is the same as the address specified in
the program that you will be using to send commands. (Use the GPIB key.)
5.
Unlock the calibration function by sending these commands:
:DIAG:KEIT:CAL:UNLOCK
:CAL:PROT:CODE ‘K1002182’
(The above command shows the default code, KI002182. Substitute the correct code if
changed.)
6.
Send the following command to initiate calibration:
:CAL:PROT:INIT
Calibration
2-13
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......