4.
Using either RANGE key, select FULL, then press ENTER. The unit will then remind
you to disconnect the input cable:
REMOVE INPUT
5.
Make sure the input cable is disconnected, then press ENTER. The unit will then
perform auto-calibration, a process that will take several minutes to complete. During
this period, the unit will display:
ACAL
NOTE
LOW-LVL ACAL performs a new gain calibration for the 10mV range only based on
10V calibration. FULL ACAL performs both a 100V and 10mV gain calibration
based on the last 10V calibration.
Performing the verification test procedures
Test summary
•
DC voltage measurement accuracy
•
Temperature measurement accuracy
•
Analog output accuracy
If the Model 2182 is not within specifications and not under warranty, see the calibration
procedures in
for information on calibrating the unit.
Test considerations
When performing the verification procedures:
•
Be sure to restore factory front panel defaults, and perform a full auto-calibration
(ACAL), as outlined above.
•
Make sure that the Model 2182 and test equipment are properly warmed up, and that the
test equipment is connected to the appropriate Model 2182 jacks.
•
Be sure the test equipment is set up for the proper function and range.
•
Allow the test signal to settle before making a measurement.
•
Use only copper-to-copper connections to minimize thermal EMFs.
•
Make sure that all connections are clean and free of oxidation to avoid thermal EMFs
that could affect measurement accuracy. Clean oxidation from connector terminals with
DeoxIT cleaning solution.
•
Keep test connections and all equipment away from drafts to avoid thermal drift.
•
Do not connect test equipment to the Model 2182 through a scanner, multiplexer, or
other switching equipment.
•
Do not use autoranging because autorange hysteresis may cause the Model 2182 to be
on an incorrect range.
Performance Verification
1-5
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......