Model 2110 5½ Digit Multimeter Reference Manual
Section 3: Basic DMM operation
2110-901-01 Rev. C/August 2013
3-53
Limits
Limit testing allows you to set high and low limit values. When the reading falls outside these limits,
the instrument beeps and displays the "HI" or "LO" message.
The limit values are stored in volatile memory. The values are reset to zero when the instrument is
turned off. Limits can be used with all measurements (except continuity and diode measurements).
For information on using the USB port as a status indicator for limit testing, see
(on page C-5).
Front-panel operation
Perform the following steps to use the limits feature:
1. Press the
SHIFT
and
NULL
(
MATH
) keys.
2.
Use the ◄ and ► keys to display
LIMITS
and press
ENTER
.
3. Set the high and low limits:
a. Use the ◄ and ► keys to display
HIGH LIMIT
and press
ENTER
.
Note that if you select the
RUN
menu option instead, the limits feature will enable using the
present limit values.
b. Use the ◄, ►, ▲, and ▼ keys to set the high limit. Note that the polarity sign can be
toggled between plus and minus. Press the
ENTER
key.
c. Use the ◄ and ► keys to display
LOW LIMIT
and press
ENTER
.
d. Use the ◄, ►, ▲, and ▼ keys to set the low limit and press
ENTER
.
4. With
RUN
displayed, press
ENTER
. The MATH indicator turns on and limit testing starts.
You can disable the math operation by pressing a different function key. You can also disable it from
the math menu as follows:
1. Press the
SHIFT
and
NULL
(
MATH
) keys.
2.
Use the ◄ and ► keys to display
OFF
and press
ENTER
.