![Keithley 2110 Reference Manual Download Page 48](http://html.mh-extra.com/html/keithley/2110/2110_reference-manual_661855048.webp)
Model 2110 5½ Digit Multimeter Reference Manual
Section 3: Basic DMM operation
2110-901-01 Rev. C/August 2013
3-5
Connections for ratio measurements
Figure 17: Connections for ratio measurements
The reference signal must be between -2.1 V and 2.1 V.
Remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following commands to select the ratio function, set a measurement range, and perform a
calculation:
FUNCtion "VOLTage[:DC]:RATio"
Select ratio function
VOLTage[:DC]:RANGe <n>
Set DCV range for the input signal
<n>
= -1000 to 1000
Expected DCV reading for the input signal
VOLTage[:DC]:RANGe:AUTO <b>
Enable or disable autorange for the input signal
<b>
= 0 or OFF
Disable autorange
= 1 or ON
Enable autorange
READ?
Perform ratio calculation and acquire reading